Operation: 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 

Top:

 DataReader program version 

Middle:

 Reset “r” number  

Bottom:

 Optics side of MicroCure 

which needs to be exposed to UV 

1.  Insert the MicroCure radiometer into the slot in the upper left side 

of the DataReader. 

2.  Press  the  RESET  button  to  activate  the  unit  from  standby/sleep 

mode and turn the unit on to active data collection mode.    

3.  The  DataReader  will  briefly  display  the  software  version  (e.g., 

P1.02 or similar)  

4.  The number of readings taken by the MicroCure is displayed as "r 

xxx" where "xxx" is the number of unit resets. Then the display will 
read "0.00" to indicate that the unit is ready to take a reading. 

5.  Remove  the  MicroCure  from  the  DataReader  and  place  it  in  the 

desired UV process location with the optics window side facing the 
UV source(s).  

 

6.  The MicroCure

 

radiometer must be exposed to UV energy within 

four minutes after being reset or it will return to a standby/sleep 
mode.  The UV irradiance level be above the start threshold level 
as described in Appendix A.  

 

7.  The MicroCure

®

 it will return to standby/sleep mode if the start 

threshold is not exceeded. Only runs above threshold are counted 
towards the limits of 200 readings. The DataReader will 
automatically shut down after thirty seconds of inactivity. 

 

8.  Remove the MicroCure

 

from the UV process environment and 

place it back into the slot in the DataReader. Press the SELECT 
button.  The DataReader will display the total UV energy in Joules 
per square centimeter (J/cm²) or the irradiance in Watts per 
square centimeter (W/cm²).   

 

9.  To toggle between total energy and irradiance, press and hold the 

SELECT button 

 

10.   To clear the current reading and to reset the MicroCure for its 

next reading, press the RESET button. 

Содержание MicroCure DataReader

Страница 1: ...icroCure Radiometer and MicroCure DataReader User Manual MicroCure Radiometer and MicroCure DataReader Electronic Instrumentation and Technology EIT LLC 309 Kelly s Ford Plaza SE Leesburg VA 20175 USA...

Страница 2: ...s 6 Operation 7 Error Messages 8 Maintenance 8 Appendix A MicroCure Product Specifications 11 Appendix B DataReader Product Specifications 12 Appendix C Regulatory Statement 13 Appendix D Technical Su...

Страница 3: ...ng applications that would ordinarily be inaccessible to larger radiometers Applications for the MicroCure include Narrow conveyor curing such as credit card processing Small container curing bottles...

Страница 4: ...is not exposed to UV above the start threshold point at which the MicroCure starts counting UV within four minutes it will enter sleep mode and not count the current wake up session as a reading o See...

Страница 5: ...ry symbol auto shut off and works with all MicroCure units Top Left Data Reader with Display Bottom Left DataReader display showing irradiance value of 1 435 W cm2 in UVA Top Right Top display shows a...

Страница 6: ...a UV source is based on the source type UV Broadband Sources are mercury based and characterized by a broad i e wide output across the UV spectrum Typical bulb types include mercury sometimes called...

Страница 7: ...s value by measuring the lamps irradiance many times during the data collection run and then summing these values together Energy Density is the total amount of energy exposure over time It is the mat...

Страница 8: ...damage both the radiometer and or your equipment Do not allow the two case sides of a MicroCure radiometer to contact each other Doing so will reset the unit and any stored data will be lost Do not a...

Страница 9: ...h the optics window side facing the UV source s 6 The MicroCure radiometer must be exposed to UV energy within four minutes after being reset or it will return to a standby sleep mode The UV irradianc...

Страница 10: ...rradiance values collected with the MicroCure should be consistent when consistent data collection techniques are followed The irradiance values displayed by the MicroCure are instantaneous peak irrad...

Страница 11: ...dip by placing a used swab back into the IPA This prevents prevent contamination of the IPA from any material picked up from the optics o Use the second clean swab in a gentle circular motion to dry t...

Страница 12: ...used in inks coatings adhesives and resins that may be near or come in contact with the MicroCure The customer assumes all responsibility for damage to their EIT instrument as a result of chemical co...

Страница 13: ...needed to activate the MicroCure Use a MicroCure with the correct operating dynamic range 10 Watt UVA or UVV 50 mW cm2 2 Watt UVA or UVV 10 15 mW cm2 200 milliWatt UVA 1 2 mW cm2 Sample Rate Smooth Mo...

Страница 14: ...ivate reset the unit Display 4 digit LCD resolution shown to 0 001 Joule Watt cm24 digit liquid crystal display Dimensions 5 77 X 4 38 x 1 45 139 7 x 108 x 36 8 mm Weight 11 75 oz 333 11 grams Operati...

Страница 15: ...326 1 2005 and the specific tests performed were Emissions EN 55011 2007 Radiated Immunity EN 61000 4 3 2006 A1 2008 and Electrostatic Discharge EN 61000 4 2 2009 The MicroCure and MicroCure DataReade...

Страница 16: ...der this warranty is limited to replacing or repairing the defective goods at our option EIT shall provide all materials and labor required to adjust repair and or replace the defective goods at no co...

Страница 17: ...te packaging if the original packaging is not used The customer is responsible for insuring the unit during transportation to EIT Once received EIT will determine if the issue will be covered under wa...

Отзывы: