Product Overview  

 

MicroCure UV Radiometer  

 

Small (1.3” x 0.95” x 0.25” /33 x 24.13 x 6.35 mm) microprocessor based, 
electro-optical  instrument  that  measures  the  total  energy  density 
(Joules/cm

2

) and  peak irradiance (Watts/cm

2

 

 

Lightweight (0.33 ounces/9.4 grams), battery powered radiometer with a fast 
sample  rate  (2048  Hz),  adapts  to  a  variety  of  demanding  physical  and 
thermal measurement environments where a larger radiometer will not fit  
 

 

Each MicroCure is good for 200 readings or one year; whichever comes first 

 

Available Bands  

o

  UVA (320-390 nm) 

o

  UVV (390-445 nm) 

 

NOTE:

 The responsivity of the EIT MicroCure UVA and UVV bands are similar but are not identical to the UVA 

and UVV bands used in other EIT instruments.     
 

 

Available Dynamic Ranges: 

o

  UVA: 10 Watt, 2 Watt or 200 milliWatt ranges 

o

  UVV: 10 Watt or 2 Watt ranges 

 

 

Suggested Operating Ranges: 

Unit Type  

 

 

Suggested Operating Range  

o

  10 Watt (UVA or UVV):    

500 mW/cm

2

 to 10 W/cm

  

o

  2  Watt (UVA or UVV):    

100 mW/cm

2

 to 2 W/cm

  

o

  200 milliWatt (UVA):       

10 mW/cm

2

 to 200 mW/cm

2

 

 

 

Once activated, the MicroCure should be exposed to UV irradiance in the Suggested Operating Range within 
four minutes.   

 

If the MicroCure is not exposed to UV above the start threshold (point at which the MicroCure starts counting 
UV) within four minutes, it will enter ‘sleep’ mode and not count the current ‘wake-up’ session as a ‘reading’ 

o

  See Appendix A for the approximate start threshold values needed to start the radiometer 

o

  For best performance, the MicroCure should be used in the Suggested Operating Range   

 

 

The MicroCure utilizes the two halves of the metal case to transfer data. Do not “short circuit” the two case halves 
on  metal.  This  includes  having  the  cases  halves  touch  metal  and/or  handling  them  with  conductive  coatings 
and/or moisture on your hands. See the section on Handling and Safety Precautions for further details. 

 
 

MicroCure DataReader   

 

The  DataReader  is  a  portable  device  used  to  communicate  download  and  display  UV  data  collected  by  the 
MicroCure radiometer.  

 

 

The “SELECT” button toggles the display between Energy Density and Irradiance values.  

 

The display shows the UV band of the MicroCure radiometer.   

 

The “RESET” button is used to “activate” a MicroCure unit and place it into “active data collection” mode.  

 

When activated, the DataReader display indicates the number of resets (r) and the count of subsequent reading 

above threshold that an individual MicroCure unit has undergone.  

 

Example: Reset value of ‘r 118’ is shown below. On the next reset, a reset value of ‘r 119’ is shown  

MicroCure showing range  

(10 Watts) and band (UVA) 

Содержание MicroCure DataReader

Страница 1: ...icroCure Radiometer and MicroCure DataReader User Manual MicroCure Radiometer and MicroCure DataReader Electronic Instrumentation and Technology EIT LLC 309 Kelly s Ford Plaza SE Leesburg VA 20175 USA...

Страница 2: ...s 6 Operation 7 Error Messages 8 Maintenance 8 Appendix A MicroCure Product Specifications 11 Appendix B DataReader Product Specifications 12 Appendix C Regulatory Statement 13 Appendix D Technical Su...

Страница 3: ...ng applications that would ordinarily be inaccessible to larger radiometers Applications for the MicroCure include Narrow conveyor curing such as credit card processing Small container curing bottles...

Страница 4: ...is not exposed to UV above the start threshold point at which the MicroCure starts counting UV within four minutes it will enter sleep mode and not count the current wake up session as a reading o See...

Страница 5: ...ry symbol auto shut off and works with all MicroCure units Top Left Data Reader with Display Bottom Left DataReader display showing irradiance value of 1 435 W cm2 in UVA Top Right Top display shows a...

Страница 6: ...a UV source is based on the source type UV Broadband Sources are mercury based and characterized by a broad i e wide output across the UV spectrum Typical bulb types include mercury sometimes called...

Страница 7: ...s value by measuring the lamps irradiance many times during the data collection run and then summing these values together Energy Density is the total amount of energy exposure over time It is the mat...

Страница 8: ...damage both the radiometer and or your equipment Do not allow the two case sides of a MicroCure radiometer to contact each other Doing so will reset the unit and any stored data will be lost Do not a...

Страница 9: ...h the optics window side facing the UV source s 6 The MicroCure radiometer must be exposed to UV energy within four minutes after being reset or it will return to a standby sleep mode The UV irradianc...

Страница 10: ...rradiance values collected with the MicroCure should be consistent when consistent data collection techniques are followed The irradiance values displayed by the MicroCure are instantaneous peak irrad...

Страница 11: ...dip by placing a used swab back into the IPA This prevents prevent contamination of the IPA from any material picked up from the optics o Use the second clean swab in a gentle circular motion to dry t...

Страница 12: ...used in inks coatings adhesives and resins that may be near or come in contact with the MicroCure The customer assumes all responsibility for damage to their EIT instrument as a result of chemical co...

Страница 13: ...needed to activate the MicroCure Use a MicroCure with the correct operating dynamic range 10 Watt UVA or UVV 50 mW cm2 2 Watt UVA or UVV 10 15 mW cm2 200 milliWatt UVA 1 2 mW cm2 Sample Rate Smooth Mo...

Страница 14: ...ivate reset the unit Display 4 digit LCD resolution shown to 0 001 Joule Watt cm24 digit liquid crystal display Dimensions 5 77 X 4 38 x 1 45 139 7 x 108 x 36 8 mm Weight 11 75 oz 333 11 grams Operati...

Страница 15: ...326 1 2005 and the specific tests performed were Emissions EN 55011 2007 Radiated Immunity EN 61000 4 3 2006 A1 2008 and Electrostatic Discharge EN 61000 4 2 2009 The MicroCure and MicroCure DataReade...

Страница 16: ...der this warranty is limited to replacing or repairing the defective goods at our option EIT shall provide all materials and labor required to adjust repair and or replace the defective goods at no co...

Страница 17: ...te packaging if the original packaging is not used The customer is responsible for insuring the unit during transportation to EIT Once received EIT will determine if the issue will be covered under wa...

Отзывы: