SM150 User Manual 1
Technical Reference
26
Technical Reference
Specifications
Accuracy
3.0% vol over 0 to 70 % vol and 0-60°C
Measurement range
0 to 100% vol but less accurate above 70%vol
3
Salinity error
5% vol over 100 to 1000 mS.m
-1
and 0-60% vol
Conductivity response
See page 27
Temperature sensitivity
See page 28
Sampling volume
See page 29
Output signal
0-1 V differential
0 to 60% nominal
Output compatible with
HH150, HH2, GP1, DL6, DL2e
Maximum cable length
1m (HH150 meter)
100m (GP1, DL6 and DL2e data loggers)
Power requirement
5-14VDC, 18mA for 1s
Operating range
-20 to +60°C
Environment
IP68
4
Sample volume
55 x 70mm diameter
Dimensions/Weight
143 x 40 mm diameter/ 0.1 kg
3
In water (no soil present) the reading may not be 100% vol. It depends on a0 and a1 but can still be
used as a quick check that the unit is working. See also page 25
4
With Delta-T supplied cables