SM150 User Manual 1
Operation
14
Logger connections and configuration
GP1 Logger
4 SM150s can connect to a GP1, but results
recorded via channels 3&4 are less accurate.
Requirements
GP1 logger (with v1.47 firmware or later)
PC running DeltaLINK (version 2.6 or later)
SM150 with SMSC/sw-05 cable
Channel 1 and 2 wiring
SM150 wire Colour GP1 terminal
Power 0V
brown
CH1
/2
(GND)
Power V+
white
CH1
/2
(PWR)
Signal HI
blue
CH1
/2
(+)
Signal LO
black
CH1
/2
(-)
Cable shield
green
CH1
/2
(GND)
Not used
grey
Not connected
Channel 3 and 4 wiring
SM150 wire Colour GP1 terminal
Power 0V
brown
CH1
/2
(GND)
or
WET GND
Power V+
white
CH1
/2
(PWR)
or
WET PWR
Signal HI
blue
Temp3
/4
(IN)
Signal LO
black
Temp3
/4
(GND)
Cable shield green
Temp3
/4
(GND)
or
WET GND
Not used
grey
Not connected
In the DeltaLINK sensor menu configure each sensor type
SM150
.
Note: channels 1&2 are wired differently to 3&4. Note also that the relative
positions of the terminals change on the left and right sides of the GP1
See also
GP1 Quick Start Guide
and the DeltaLINK on-line Help.