8
Maintenance
Page 38 / 52
to ensure that the processing, as actually performed using
equipment, materials and personnel in the processing facility,
achieves the desired result. This requires verification and/or
validation and routine monitoring of the process. Likewise, any
deviation by the processor from the instructions provided should be
properly evaluated for effectiveness and potential adverse
consequences.
8.4
Disinfection Components
Disinfection components
Motor Handpiece
Handpiece Base
Adapter
Wipe all the surfaces with a cloth lightly moistened with Ethanol for
Disinfection (Ethanol 70 to 80vol%) at least 2min, repeat for 5 times.
NOTE
1. Do not use disinfectants other than alcohol for disinfection.
2. Do not use excessive alcohol to prevent alcohol from seeping
Into the parts and damaging the internal parts.
3. Disinfect before and after each use.