![Eighteeth E-xtreme Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/eighteeth/e-xtreme/e-xtreme_user-manual_2369378034.webp)
8
Maintenance
Page 34 / 52
sing area to avoid any damage and contaminate
on to the environment.
Preparation for
Decontaminati-
on
The devices must be reprocessed in a disassemb-
led state.
WARNING
1. Do not fail to take out the file before cleaning
the contra angle.
2. Observe suitable personal protective measure-
es.
Pre-Cleaning
Do a manual pre-cleaning, until the components
are visually clean. Submerge the components in a
cleaning solution and flush the lumens with a
water jet pistol with cold tap water for at least 10
seconds. Clean the surfaces with a soft bristol
brush.
Cleaning
Regarding cleaning/disinfection, rinsing and
drying, it is to distinguish between manual and
automated reprocessing methods. Preference is
to be given to automated reprocessing methods,
especially due to the better standardizing
potential and industrial safety.
Automated Cleaning:
Carefully put the components into the
washer-disinfector on a tray and set the
parameters as follows, then start the program:
·
4 min pre-washing with cold water (<40°C);
·
emptying
·
5 min washing with a mild alkaline cleaner
at 55°C;
·
emptying
·
3 min neutralising with warm water (>40°C);
·
emptying