![EIC transfusio-therm 2000 Скачать руководство пользователя страница 6](http://html.mh-extra.com/html/eic/transfusio-therm-2000/transfusio-therm-2000_quick-reference-manual_2369143006.webp)
!" # $
"
% "
& '
(
' )
Meaning of the failure codes safety shutdown
„transfusio-therm 2000“ controller version 7 and 8
Code
Error
0
Door switch, different signals
Equipment protective cut-off
1
Error turning / blocked or protective
cut-off
2nd system
Equipment protective cut-off
2
Error turning / over-wound
Equipment protective cut-off
3
Error magnetron / does not switch on or
protective switching
2nd system
Equipment protective cut-off
4
Error magnetron / does not switch off
Equipment protective cut-off
5
Error antenna / turns too slowly
Equipment protective cut-off
6
Deficient self-test
Equipment protective cut-off
7
T1...T6 > 41
o
C
Equipment protective cut-off
8
Full sensor 1 open,
Full sensor 2 closed
Equipment protective cut-off
Intake 1
9
Full sensor 2 open,
Full sensor 1 closed
Equipment protective cut-off
Intake 1
10
Temperature difference > 1
o
C
Safety cut-off
Intake 1
11
Break in the temperature sensor cable
Safety cut-off
Intake 1
12
T1 or T2 > 39
o
C
Safety cut-off
Intake 1
13
Intake leaky (HF)
Safety cut-off
Intake 1
14
15
16
Full sensor 1 open,
Full sensor 2 closed
Protective cut-off
Intake 2
17
Full sensor 2 open,
Full sensor 1 closed
Protective cut-off
Intake 2
18
Temperature difference > 1
o
C
Safety cut-off
Intake 2
19
Break in the temperature sensor cable
Safety cut-off
Intake 2
20
T1 or T2 > 39
o
C
Safety cut-off
Intake 2
21
Intake leaky (HF)
Safety cut-off
Intake 2
22
23
24
Full sensor 1 open,
Full sensor 2 closed
Protective cut-off
Intake 3
25
Full sensor 1 closed,
Full sensor 2 open
Protective cut-off
Intake 3
26
Temperature difference >
1
o
C
Safety cut-off
Intake 3
27
Break in the temperature sensor cable
Safety cut-off
Intake 3
28
T1 or T2 > 39
o
C
Safety cut-off
Intake 3
29
Intake leaky (HF)
Safety cut-off
Intake 3
30
31