EIC transfusio-therm 2000 Скачать руководство пользователя страница 3

transfusio-therm

®

 2000      

 

 

Quick Reference Guide 

 

Status: June 2013     Page 3 of 5 

08 

Target temperature 
reached 

Automatic acoustic signal and green 
flashing LED on the compartment 
concerned 

ON 
x °C / CONFIRM 
DEFROSTED / 
CONFIRM 

The set target temperature will be 
displayed. 

09 

Confirming the 
removal 

Press the confirmation button of the 
compartment concerned. 

ON 
REMOVE BAG 

 

10 

Opening the door 

Use the door handle to open the 
door. 

ON 
LOADED 

 

11 

Removing the bag 

Follow the steps below to remove 
the bag:  

 

Raise the spring plate of the 
compartment. 

 

Remove the bag. 

 

Release the spring plate. 

ON 
--------- 

Once the bag has been removed, 
the compartment can be loaded 
again. To do so, start with Item 03. 

12 

Closing the door 

Use the door handle to close the 
door. 

ON 
VACANT 

The appliance is ready for use. For 
longer periods of non-use, turn 
the appliance off, using the ON / 
OFF power switch. 

 
III.  Safety Comments 

 

Item 

No. 

Subject 

Information 

Display 

Important: Be Sure to Pay 

Attention to the Comments Below. 

 

 

 

 

 

13 

Special operating 
note 

If a bag has been inserted into 
the compartment, with no 
operating action taking place 
(with the door open or closed), 
although a relevant prompt has 
appeared in the display, an 
acoustic signal sounded after 
approximately 5 minutes will 
remind the operator to resume 
work on the appliance.  

 

Whenever an acoustic signal is 
sounded, it is imperative that you 
go to the appliance and act upon 
the prompt. 

14 

Safety shutdown 

If the two-channel monitoring 
system of the appliance detects 
an error in any compartment 
area, the compartment will be 
blocked, as signaled by a steady 
red light of the LED.  

ON / SERVICE 
BLOCKED 

Do not use the blocked 
compartment.  
 
Proceed as described in Item 16, 
which addresses a protective 
shutdown.  

15 

Inadvertent 
loading of a 
compartment 
following a safety 
shutdown (13) 

If a shutdown compartment is 
loaded inadvertently, the red 
LED will change from a steady 
light to a flashing one. 

ON / SERVICE 
REMOVE BAG 

After pressing the confirmation 
button and removing the bag from 
the blocked compartment, you can 
continue using those 
compartments of the appliance 
which are not blocked. 

16 

Protective 
shutdown 

All the compartments are shut 
down whenever the two-channel 
monitoring system of the 
appliance detects a malfunction 
in the overall system. The 
compartment LEDs and the 
alarm LED will show steady red 
light.  

OFF 
PROTECTIVE 
SHUTDOWN 
NOTIFY SERVICE 
 

In the event of a protective 
shutdown, please be sure to turn 
the appliance off, using the ON / 
OFF power switch.   
 
Open the door and check that all 
the bags are inside the 
compartments and that they are 
not damaged. If necessary, 
remove all the bags from the 
compartments. Check the 
temperature of the bags. Destroy 

Содержание transfusio-therm 2000

Страница 1: ...suggestions when putting the transfusio therm 2000 into operation and when using it I Overview Compartment No 1 Compartment No 3 Compartment No 2 ON OFF Power Switch Start Button with LED LED Alarm Signal Confirmation Buttons with LED Compartment No 1 Compartment No 2 Compartment No 3 Operator Prompting Display Status Line User Interface for Compartment No 1 User Interface for Compartment No 2 Use...

Страница 2: ...ion the bag is to be inserted with the connectors located on the right hand side It is possible to thaw a blood bags inserted into protective bags provided the above instructions are observed 04 Closing the door Use the door handle to close the door ON 05 Confirming loading When prompted by the yellow flashing LED and the text appearing in the display press the confirmation button of the compartme...

Страница 3: ...mind the operator to resume work on the appliance Whenever an acoustic signal is sounded it is imperative that you go to the appliance and act upon the prompt 14 Safety shutdown If the two channel monitoring system of the appliance detects an error in any compartment area the compartment will be blocked as signaled by a steady red light of the LED ON SERVICE BLOCKED Do not use the blocked compartm...

Страница 4: ...with particular care when cleaning temperature measuring points silver plates After cleaning wipe all parts dry and screw the spring plates on again tightly Push the complete rotating unit from the front onto the drive shaft as far as it can go Then reinsert the safety hook Before putting the appliance into use again check that the rotating unit is properly fitted Be sure to clean the motor shaft ...

Страница 5: ...rrors all the way to error no 31 To get the number of bags heated press button no 2 Record the read out error code together with the serial number of the appliance see nameplate attached on the right hand side of the appliance in a fault report form and fax or e mail this form to EIC Ltd EIC Umwelt und Medizintechnik Ltd Rudolf Diesel Strasse 5 D 37308 Heilbad Heiligenstadt Germany Telephone 49 36...

Страница 6: ...quipment protective cut off Intake 1 10 Temperature difference 1 o C Safety cut off Intake 1 11 Break in the temperature sensor cable Safety cut off Intake 1 12 T1 or T2 39 o C Safety cut off Intake 1 13 Intake leaky HF Safety cut off Intake 1 14 15 16 Full sensor 1 open Full sensor 2 closed Protective cut off Intake 2 17 Full sensor 2 open Full sensor 1 closed Protective cut off Intake 2 18 Tempe...

Отзывы: