EIC transfusio-therm 2000 Скачать руководство пользователя страница 1

transfusio-therm

®

 2000      

 

 

Quick Reference Guide 

 

Status: June 2013     Page 1 of 5 

The  Quick  Reference  Guide  must  be  kept  directly  on  the  transfusio-therm

®

  2000  in  a 

clearly  visible  manner.  Be  certain  to  observe  the  following  sequence  of  operations  and 
suggestions when putting the transfusio-therm

®

 2000 into operation and when using it. 

 

I. 

Overview  

 
 
 
 
 

 

Compartment No. 1 

Compartment No. 3 

Compartment No. 2 

ON / OFF Power Switch 

Start Button  
with LED 

LED  
Alarm Signal 

Confirmation Buttons with LED 
Compartment No. 1  
Compartment No. 2  
Compartment No. 3 

Operator-Prompting Display 

Status Line 
User Interface for Compartment No. 1 
User Interface for Compartment No. 2 
User Interface for Compartment No. 3 

Nameplate

 

Содержание transfusio-therm 2000

Страница 1: ...suggestions when putting the transfusio therm 2000 into operation and when using it I Overview Compartment No 1 Compartment No 3 Compartment No 2 ON OFF Power Switch Start Button with LED LED Alarm Signal Confirmation Buttons with LED Compartment No 1 Compartment No 2 Compartment No 3 Operator Prompting Display Status Line User Interface for Compartment No 1 User Interface for Compartment No 2 Use...

Страница 2: ...ion the bag is to be inserted with the connectors located on the right hand side It is possible to thaw a blood bags inserted into protective bags provided the above instructions are observed 04 Closing the door Use the door handle to close the door ON 05 Confirming loading When prompted by the yellow flashing LED and the text appearing in the display press the confirmation button of the compartme...

Страница 3: ...mind the operator to resume work on the appliance Whenever an acoustic signal is sounded it is imperative that you go to the appliance and act upon the prompt 14 Safety shutdown If the two channel monitoring system of the appliance detects an error in any compartment area the compartment will be blocked as signaled by a steady red light of the LED ON SERVICE BLOCKED Do not use the blocked compartm...

Страница 4: ...with particular care when cleaning temperature measuring points silver plates After cleaning wipe all parts dry and screw the spring plates on again tightly Push the complete rotating unit from the front onto the drive shaft as far as it can go Then reinsert the safety hook Before putting the appliance into use again check that the rotating unit is properly fitted Be sure to clean the motor shaft ...

Страница 5: ...rrors all the way to error no 31 To get the number of bags heated press button no 2 Record the read out error code together with the serial number of the appliance see nameplate attached on the right hand side of the appliance in a fault report form and fax or e mail this form to EIC Ltd EIC Umwelt und Medizintechnik Ltd Rudolf Diesel Strasse 5 D 37308 Heilbad Heiligenstadt Germany Telephone 49 36...

Страница 6: ...quipment protective cut off Intake 1 10 Temperature difference 1 o C Safety cut off Intake 1 11 Break in the temperature sensor cable Safety cut off Intake 1 12 T1 or T2 39 o C Safety cut off Intake 1 13 Intake leaky HF Safety cut off Intake 1 14 15 16 Full sensor 1 open Full sensor 2 closed Protective cut off Intake 2 17 Full sensor 2 open Full sensor 1 closed Protective cut off Intake 2 18 Tempe...