P:
206.402.5241
E:
[email protected] | 169 Western Ave W, Seattle, WA 98119 | www.eagleharbortech.com
Page 8 of 32
3.2
Operational Space
The NSP voltage, pulse width, and pulse repetition frequency are all individually adjustable. For the load
types indicated, the following figures show the accessible operating space.
•
The minimum pulse width is measured at maximum output voltage, defined as the full width at
half-maximum (FWHM). Shorter pulses are possible at voltages below the NSP’s maximum
output voltage.
•
Simultaneous maximum for each specification is limited by both available power and by the load
driven.
•
If the user sets the operating parameters to a region outside the operating area, the NSP will self-
protect by dropping the output voltage.
Figure 2.
Available Operating Space for the NSP-120-20-P-500-H