EFSEN UV BARx1 Скачать руководство пользователя страница 8

Product: 
Project: 

UV BARx1 version 2.2 
2201075 

Page: 

 

Latest updated: 

7 af 12 
30.07.2020 

 

www.efsen.dk 

 

 

3.

 

End of use 

 

a.

 

When finished with use of UV BARx1 turn OFF by pressing 

the round button. 

 

 

b.

 

DO NEVER LEAVE UV BARx1 TURNED ON!

 

 

 

 

 

c.

 

Remove the cable from the wall plug so electricity is disconnected.

 

d.

 

Now, it is safe to remove face shield, gloves and protective clothing.

 

e.

 

Ensure that the UV lamp has no damages. 

 

 

f.

 

Transport the UV BARx1 in a protected way so that the UV 

lamp inside is not damaged.  

Do a visual check for damages after each transport. 

 

 

 

 

 

Содержание UV BARx1

Страница 1: ...Safety and User Manual Version 2 2 ENGLISH ...

Страница 2: ...Manual Version 2 2 Table of contents Safety and Health 2 Please watch the Safety Instruction User Guide video 4 Clothing and safety for the use of UV BARx1 5 Operation How to use the UV BARx1 6 Maintenance 8 Disclaimer Safety 10 Disclaimer Process guaranty 10 Product Warranty 10 EU Declaration of Conformity 11 ...

Страница 3: ... make the materials more brittle give them a chalky surface and fade or change the color and other effects UV BARx1 is emitting UV C light in 254nm which inactivates by changing the RNA and prevents the reproduction of bacteria viruses and protozoa so normal cleaning is still needed to remove dirt and dust To secure safe and proper operation the system must be handled and operated correctly and us...

Страница 4: ...ake precautions and be careful to avoid burns Electrical voltage The supply voltage for the UV lamps is 230VAC When installing the UV lamp be sure to place the connection cables in a secure position so that the insulation around the cables cannot be damaged Warnings regarding voltage Service and repair must only be performed by a certified electrician Disconnect the power cord from the wall plug b...

Страница 5: ...s that are available to them Please watch the Safety Instruction User Guide video It is important that you watch and understand our video guide We will teach the operator step by step how to use the UV BARx1 correctly and safely Please follow this link www efsen dk products uv disinfection uv barxone safety and user guide or scan the QR code Please be aware that UV deactivation of virus and bacter...

Страница 6: ...ands must be protected by light proof gloves Verify that all skin areas are covered by the gloves or by the sleeves c Skin must be covered and protected with light proof clothes Also remember to cover your neck d Shoes must be worn Sandals or bare feet are not allowed Also here all skin must be covered 2 Safe and proper use a Make sure that all humans within a distance of 10 m from UV BARx1 the so...

Страница 7: ...nd activate the UV BARx1 by pressing the ON button c The UV lamp will turn on immediately and will be ready to use 2 Using the UV BARx1 a Hold UV BARx1 with power cord facing the operator and the light surface pointing away from operator b Keep UV BARx1 in 50 100mm distance of object surface to be treated c Pan UV BARx1 over objects surfaces at 20 cm s this corresponds to using 3 seconds on a keyb...

Страница 8: ...the round button b DO NEVER LEAVE UV BARx1 TURNED ON c Remove the cable from the wall plug so electricity is disconnected d Now it is safe to remove face shield gloves and protective clothing e Ensure that the UV lamp has no damages f Transport the UV BARx1 in a protected way so that the UV lamp inside is not damaged Do a visual check for damages after each transport ...

Страница 9: ... cloths and soap 2 Changing the UV lamp a The UV BARx1 must be turned off for 2 minutes for cool down b Disconnect UV BARx1 from wall plug before changing the UV lamp c Use gloves do not touch UV lamps with your hands and fingers d Remove the protective screen in the bottom of the UV BARx1 using a screwdriver e Release the UV lamp in the end where the plug is by pressing the red tab and pulling ge...

Страница 10: ...small portion of Mercury so they must be disposed accord to the local legislation for example at the local recycling station 4 Broken UV lamps Damaged UV lamps a If an accident happens and the UV lamp breaks then do not touch glass or any other residue with bare hands Be aware that UV lamps contains mercury or amalgam b Make sure to have ventilation in space open window or door c Gently push bigge...

Страница 11: ...his appliance is not intended for use by persons including children with reduced physical sensory or mental capabilities or lack of experience and knowledge unless they have been given supervision or instruction concerning use of the appliance by a person responsible for their safety Children should be supervised to ensure that they do not play with the appliance Disclaimer Process guaranty EFSEN ...

Страница 12: ...09 125 EC The Ecodesign Directive EC No 245 2009 Commission Regulation implementing ecodesign requirements for fluorescent lamps without integrated ballast for high intensity discharge lamps and for ballasts and luminaires able to operate such lamps EU No 1194 2012 Commission Regulation implementing ecodesign requirements for directional lamps light emitting diode lamps and related equipment EU No...

Страница 13: ...ntrol gear for fluorescent lamps EN 55015 2013 Limits and methods of measurement of radio disturbance characteristics of electrical lighting and similar equipment EN 61547 2009 Equipment for general lighting purposes EMC immunity requirements EN 61000 3 2 2014 Electromagnetic compatibility EMC Part 3 2 Limits Limits for harmonic current emissions equipment input current 16 A per phase EN 61000 3 3...