Technical data
A732-01-880 Issue A
2
Technical data
2.1
Operating and storage conditions
2.2
Performance
2.2.1
General
Note:
If the pump is operated outside the specified limits, then the inverter may reduce the motor speed; tip seal
wear rate will be increased.
Table 1 - Operating and storage conditions
Operating and storage conditions
XDS100B
Ambient temperature range (operation)
+10 °C to +40 °C
Maximum surface temperature of pump body under normal operating conditions
and maximum ambient temperature
+40 °C to +70 °C
Maximum humidity (operation)
90% RH
Ambient temperature range (storage)
–30 °C to +70 °C
Table 2 - CSA approval environmental conditions
Parameter
XDS100B
Pollution
Pollution degree 2
Installation
Installation category II
Altitude restriction
Max 2000 m
Area of use
Indoor use
Table 3 - General characteristics
General
XDS100B
Nominal peak pumping speed
95 m
3
/h
*
*
XDS100B used in combination with XDS35i backing pump.
Maximum permitted continuous inlet pressure
20 mbar
Maximum permitted continuous exhaust pressure
60 mbar
Maximum permitted inlet pressure for short duration
1.5 bar(a)
Maximum chamber volume to pump down from atmospheric pressure
100 litres
Maximum chamber volume for cyclic duty - maximum 6 cycles per hour
50 litres
Suckback protection
Provided by backing pump
Leak tightness
<1x10
-6
mbar l s
-1