A746‐03‐885A ‐ Maintenance
Pump does not achieve
ultimate total pressure
or normal pumping
speed.
Centring ring not correctly
positioned or leak in the pipeline
or vacuum system?
Check pump with a vacuum
gauge directly at pump inlet
port, check connections and
line.
Long narrow line?
Use line with larger diameter,
length as short as possible.
Pump has been exposed to
condensate?
Run pump at atmospheric
pressure for a few minutes.
Deposits have been formed
inside the pump?
Clean and inspect pump heads.
Valves or diaphragm damaged?
Replace valves and/or
diaphragms.
Outgasing substances or vapour
generated in the process?
Check process parameters.
Only 24 V DC version:
Pump temperature too high
(reduced motor speed)?
Ensure sufficient cooling of the
pump or reduce inlet pressure.
Pump too noisy.
Atmospheric or high pressure at
inlet port?
Connect hose to pump outlet.
Diaphragm clamping disc loose? Perform maintenance.
None of above mentioned
causes?
Contact your supplier.
Pump seized.
Contact your supplier.
Table 9
Troubleshooting (continued)
Fault
Possible cause
Remedy
Содержание XDD1 115 V AC
Страница 8: ...This page has been intentionally left blank Page iv A746 02 885A List of Tables ...
Страница 16: ...Page 8 A746 03 885A Technical data Figure 3 Dimensional drawing XDD 1 dual voltage motor ...
Страница 17: ...Page 9 A746 03 885A Technical data Figure 4 Dimensional drawing XDD 1 24 V DC ...
Страница 44: ...This page has been intentionally left blank Page 36 A746 03 885A Spares and accessories ...
Страница 45: ...This page has been intentionally left blank ...
Страница 46: ...This page has been intentionally left blank ...
Страница 48: ...P800 80 000 Issue T edwardsvacuum com ...