D0300041001EN - Computer interface
Page 41
RVN (Read Version Number)
Reading the version number of the instrument software.
Example:
RVN<CR>
Response:
x.xx<CR>
RSS (Read Setpoint Status)
Reading the status of the switching thresholds for a channel.
Example:
RSS[a]<CR>
Response:
b[,][TAB]c<CR>
SHV (Set HV on/off)
Setting HV on/off for IG, if switch-on and switch-off type have been set to manual.
Example:
SHV[a, b]<CR>
Response:
OK<CR>
Parameter
Description
x.xx
Version number
Parameter
Description
a
Channel number
1 = Channel 1 (PRG 1)
2 = Channel 2 (PRG 2)
3 = Channel 3 (IG)
b
Status SP1
0 = low
1 = high
c
Status SP2
0 = low
1 = high
Parameter
Description
a
Channel number
3 = Channel 3 (IG)
b
HV on/off
0 = off
1 = on