D0300041001EN - Computer interface
Page 40
RPV (Read Pressure Value)
Reading a pressure value for a measurement channel.
Example:
RPV[a]<CR>
Response:
b[,][TAB]x.xxxxE±xx
SSC
Set Sensor Control
Set type of sensor control for a measurement channel.
RSP
Read Setpoint
Read switching thresholds for a channel.
SSP
Set Setpoint
Set switching thresholds for a channel.
Note:
All changes to values are only permanently saved to the EEPROM after having issued the write
command SAC. For this reason, after having set all values, save the current configuration
through the SAC write command before restarting the instrument.
Parameter
Description
a
Channel number
1 = Channel 1 (PRG 1)
2 = Channel 2 (PRG 2)
3 = Channel 3 (IG)
b
Status
0 = Measured value OK
1 = Measured value < measurement range
2 = Measured value > measurement range
3 = Measured values << measurement range (
)
4 = Measured values >> measurement range (
)
5 = Sensor off (
)
6 = HV on (
)
7 = Sensor error (
)
9 = No sensor(
)
10 = No switch-on or switch-off threshold (
)
12 = Pirani error (
)
16 = Measured value OK degas function
x.xxxxE±xx
Pressure value of the selected channel in the current unit
of measurement
Table 24 Mnemonics for read and write commands (continued)
Read and write commands
Description
Содержание PGC202
Страница 10: ...This page has been intentionally left blank D0300041001EN Contents Page vi ...
Страница 14: ...D0300041001EN Safety Page 4 This page has been intentionally left blank ...
Страница 38: ...D0300041001EN Operation Page 28 This page has been intentionally left blank ...
Страница 46: ...D0300041001EN Parameters Page 36 This page has been intentionally left blank ...
Страница 58: ...D0300041001EN Computer interface Page 48 This page has been intentionally left blank ...
Страница 62: ...D0300041001EN Storing and waste disposal Page 52 This page has been intentionally left blank ...
Страница 64: ...P800 80 000 Issue T edwardsvacuum com ...