D0300040001EN - Parameters
Page 32
Type of gas correction CPG (
)
The CPG gauge heads are normally calibrated for measurements in nitrogen or air. With the
aid of the parameter
you can set up the measurement channel to other types of gas.
For this proceed as follows:
Select the parameter
.
Press the pushbutton (PARA).
The setup characteristic is displayed.
Use the arrow pushbuttons to change the characteristic.
The value of the parameter is changed.
You can set up the correction factor for a gauge head in the range of 0.2 - 1.0 - 8.0. A setting
of 1.0 returns the uncorrected value.
Sensor switch-on type (
)
This parameter defines how the CPG gauge head is switched on.
You can set the switch-on time to the following values:
Sensor switch-on value (
)
This parameter will only appear when the sensor switch-on time has been set to
or
(see
Through the parameter t-on you can define a switch-on value. When the pressure in the
respective measurement channel drops below the switch-on value, the gauge head is
switched on.
Sensor switch-off type (
)
This parameter defines how the CPG gauge head is switched off.
You can set up the switch-off type to the following values:
Manual. The gauge head can be switched on by pressing the pushbutton (UP)
(see
Externally through optocoupler (PLC compatible high-level > 13 V d.c.; 7 mA at
24 V d.c.).
Warm start. The sensor is switched on automatically upon switching on the
instrument. After a power failure the measurement is started automatically.
Through measurement channel PRG 1. With the aid of the next parameter t-on
the operator can define a switch-on value. When the pressure in the
measurement channel PRG 1 drops below the switch-on value, the sensor is
switched on.
Through measurement channel PRG 2. With the aid of the next parameter
the operator can define a switch-on value. When the pressure in the
measurement channel PRG 2 drops below the switch-on value, the sensor is
switched on.
Содержание PGC201
Страница 10: ...This page has been intentionally left blank D0300040001EN Contents Page vi ...
Страница 14: ...D0300040001EN Safety Page 4 This page has been intentionally left blank ...
Страница 30: ...D0300040001EN Installation Page 20 This page has been intentionally left blank ...
Страница 38: ...D0300040001EN Operation Page 28 This page has been intentionally left blank ...
Страница 46: ...D0300040001EN Parameters Page 36 This page has been intentionally left blank ...
Страница 58: ...D0300040001EN Computer interface Page 48 This page has been intentionally left blank ...
Страница 62: ...D0300040001EN Storing and waste disposal Page 52 This page has been intentionally left blank ...
Страница 63: ...This page has been intentionally left blank ...
Страница 64: ...This page has been intentionally left blank ...
Страница 67: ...This page has been intentionally left blank ...
Страница 68: ...P800 80 000 Issue T edwardsvacuum com ...