Appendix A2
M561-00-880 Issue C
Notes:
1. Some of these warnings apply to pump sensors that are only present on certain system variants.
2. Some of these warnings apply to accessories and may only be encountered if the particular accessory is
fitted.
Warning 70.13
Sensor missing
-
The dry pump temperature sensor may
have become disconnected or failed.
- Check that the sensor is
fitted and is correctly
connected.
Warning 71.13
AC Sup Missing
See manual
The pump controller cannot communicate
with the accessory module
- Check that the accessory
module is correctly
connected to the pump and
then cycle the power to the
pump.
- If this warning persists,
contact Edwards.
Warning 84.11
Exh Temp High
See manual
The exhaust pipe temperature is too high
Contact Edwards
Warning 84.13
Sensor missing
-
The exhaust pipe temperature sensor may
have become disconnected or failed.
Contact Edwards
Warning 152.01
Valve Not Shut
Check Gate Valve
The foreline gate valve has failed to close
Check the wiring and air
supply to the Gate valve.
Note:
This warning will
only clear when the
valve has
successfully been
opened and closed.
Warning 153.01
Valve Not Open
Check Gate Valve
The foreline gate valve has failed to open
Check the wiring and air
supply to the Gate valve.
Note:
This warning will
only clear when the
valve has
successfully been
closed and opened.
Warning 176.01
MB INV xxxx yyyy
xxxx yyyy
aaaaaaaaaaaaaaaa
Booster Inverter has raised a warning code. Refer to
A2.3
for more
information.
Warning 176.13
No MB Inv Comms
See manual
The pump controller cannot communicate
with the booster inverter.
Check the wiring between
the pump controller and the
booster inverter.
Warning 186.01
DP INV xxxx yyyy
xxxx yyyy
aaaaaaaaaaaaaaaa
Dry pump Inverter has raised a warning
code.
Refer to
A2.3
for more
information.
Warning 186.13
No DP Inv Comms
See manual
The pump controller cannot communicate
with the dry pump inverter.
Check the wiring between
the pump controller and the
dry pump inverter.
Warning 196.01
DP Inv xxxx yyyy
Diag aaaa bbbb
/zzzzzzzzzzzzzzz
DP 2nd Inverter has raised warning code.
Refer to
A2.3
for more
information.
Warning 196.01
DP Not Running
See manual
DP 2nd Inverter will not start up when
requested.
Check Inverter Fault History.
Warning 196.13
No DP Inv Comms
See manual
The pump controller cannot communicate
with the dry pump 2nd inverter.
Check the wiring between
the pump controller and the
dry pump inverter.
Warning 314.11
Speed Too Low
Stop Pump/Contact
Service
The pump rotational speed is at least 30%
less than demand for more than 2 minutes.
Stop the pump and perform
mechanical check of
rotation and beearings.
Table A15 - Warnings (continued)
Warning Message
on PDT
'Action' Message on
PDT
Possible Cause
Action