background image

Appendix 1 

Cell Constants 

Conductivity cells with different cell constants can be used to 

achieve greater accuracy or used to make difficult measurements 

easier. Selection of the correct cell constant is dependent on the 

conductivity range of the sample. Conductivities of various 

waters and common solutions, together with the most suitable 

cell constants are given below. 

K=0.1/cm 

For measurement of solutions with very low conductivity  e.g.: 

pure water, de-mineralise water; distilled water; boiler feed 

water 

K=1.0/cm 

For measurement of solutions with medium conductivity e.g.: 

Surface water; waste water; diluted salt solutions; fertilizers; 

electroplating rinses. 

K=10/cm 

For measurement of solutions with high conductivity e.g.: 

Strong acid; strong alkali; strong salt solutions; sea water 

Cell Conversion Table 

c.i,.s. Units

SI Units 

K�I.O/cm 

K�J00.0/m 

K�0.1/cm 

K�JO/m 

K�J0.0/cm 

K�JOOO/m 

FE

2

80 

- P

o

r

t

able C

o

nd

u

ct

i

vi

t

y Me

t

e

r - Users Manual

Содержание FE280

Страница 1: ...g the battery Instrument test Section 2 5 Calibration Measurement Section 3 6 Troubleshooting Section 4 7 Specification Section 5 8 Conductivity Cells Appendix 1 9 Cell Constants Appendix 2 10 Calibra...

Страница 2: ...hen you receive the shipment inspect the container for any signs ofdamage Note any evidence ofrough handling in transit Imme diately report any damage to the agent Note The carrier will not honour any...

Страница 3: ...9V battery The BAT flag appears on the display to indicate a low battery To install or replace the batte1y slide offthe back cover Remove the old battery and inse1i a new one ensuring that the polari...

Страница 4: ...trol by inse11ing a screwdriver into the small hole on the side ofthe instrument Turn the adjustment screw until the display reads l 000mS cm and remainsstable Select 20mS cm Select 200mS cm Select 20...

Страница 5: ...for the reading to stabilise Gentle stirring may be necessary to dislodge trapped air bubbles 4 When the reading is stable adjust the value to that ofthe standard using a screwdriver to tum the calib...

Страница 6: ...t or lack ofstiITing Cell contaminated En atic drifting readings or Return meter for servicing display reads on left hand side when calibration plug is attached In the event ofa malfunction it is impo...

Страница 7: ...compensation range Reference Temperature Temperature Coefficient Display Power Instrument Size Instrument Weight 7 0 200 S cm 0 1 0 2mS cm 1 0 20mS cm 10 0 200mS cm 100 0 1 ofreading 0 50 C 2s0 c 2 1...

Страница 8: ...xy K l0 00mS Graphite Industrial 25mm 2000mS plates Field A5010 EpoxyK l I00 s Graphite Industrial 25mm JOOmS plates Field A5011 EpoxyK 1 0 011 S Graphite Industrial 25mm 200mS plates Field A5019 Epox...

Страница 9: ...en below K 0 1 cm For measurement ofsolutions with very low conductivity e g pure water de mineralise water distilled water boiler feed water K 1 0 cm For measurement ofsolutions with medium conductiv...

Страница 10: ...2 133 2 34 1667 23 1359 135 9 35 24 1386 138 6 36 12 88mS cm 25 C 0 lM KCI oc Siem mSim oc Siem 5 8220 822 25 12880 10 9330 933 26 13130 15 10480 1048 27 3370 16 10720 1072 28 13620 17 10950 1095 29 1...

Страница 11: ...34 3 256 23 2 659 265 9 35 3 312 24 2 712 271 2 36 3 368 2 l l l 8rnS crn 25 C 0 02M KC oc mS cm mS m oc mS cm 5 74 14 7 41 25 111 80 10 83 19 8 32 26 113 77 15 92 52 9 25 27 115 74 16 94 41 9 44 28...

Отзывы: