Section  7 

Concentration 

Please refer to page 9 before proceeding. 

The following can only be carried out if the instrument has not 

been calibrated in Conductivity Mode. 

Calibration and Measurement 

Prepare up to four calibration solutions, each with a known 

© 

concentration 

Connect cell and select concentration mode 

Place cell in first standard and wait for a stable reading. The 

./J. key can be used to adjust the value to the correct decade 

of concentration. The exact value may then be entered by 

adjusting the reading using the ./J. fr keys,  (CAL flag will 

flash). And then pressing  Enter. (CAL flag will stop flash­

ing). 

Repeat © for each standard. 

Place cell in sample and read concentration. 

The calibration points may be inspected at any time as follows: 

G)

Press CHK

(JJ 

First a number identifier is displayed and then the standard 

value. 

This is repeated for each standard. 

© 

Return to the function you require by pressing the Mode 

key. 

To Back off the background conductivity 

Follow points Oto 0 

Place cell in background sample and enter 0.00 

Place cell in sample and read concentration 

13 

,,,,,,,,,� 

____ 

,, , 

Section 8 

Use of the Recorder Output 

Refer to the recorder instructions 

Connect the recorder via the red and black 4mm sockets on 

the back panel. (Red positive, Black negative) 

Ensure that the recorder is set for the appropriate range, i.e. 

Mode 

Range 

Display 

Recorder 

Cond 

0-200  l000µS/cm  l00mV

TDS 

0-200 

666mg/L

66.6mV 

Res 

0-200 

IO.OMO.cm  10.0mV

Sal 

0-200 

35.0

35.0mV 

oc 

±200 

25.0

°

25.0mV 

Range Hold 

The auto-ranging facility on the RE3 88Tx selects the most appro­

priate range for the current reading automatically. 

To disable the auto- ranging facility 

Place cell in the sample and select the required mode. 

Press the Send key followed by the Enter key. 

© 

The range currently in use is now held. 

To disable range hold press Clear key 

Exit Hold facility by pressing the Send key. 

14 

Содержание RE388TX

Страница 1: ...2 Salinity Section 6 12 Temperature Section 7 13 Concentration Section 8 14 The Recorder Output Section 9 15 Use with Printer or Computer Section 10 17 Platinizing Section 11 18 Troubleshooting Sectio...

Страница 2: ...have any ques tions about the shipment please call EDT Direct ION or your agent When you receive the shipment inspect the container for any signs ofdamage Note any evidence ofrough handling in transi...

Страница 3: ...cover J Battery Snap Figure 1 Bottom view ofinstrument showingpower connections 3 Instrument Test Procedure Refer to Figure 2 Switch the instrument on using the black switch on the back panel The mete...

Страница 4: ...vity l0MQ cm 0 MQ m Changing the multiple 0 Press the K key to display the multiple e Set the preferred units by pressing the K key to select either m orcm e Exit with the correct units set by pressin...

Страница 5: ...ESS THE Enter KEY WHEN SETTING THE CELL CONSTANT 7 Setting the Temperature Coefficient The conductivity ofelectrolytes changes with temperature The rate ofthis change is termed the temperature coeffic...

Страница 6: ...on 2 Conductivity Please refer topage 9 beforeproceeding The following can only be carried out ifthe instrument has not been calibrated in Concentration Mode Most conductivity cells are linear over a...

Страница 7: ...ding and record result noting units 11 Section 5 Salinity Salinity Measurement Please refer topage 9 beforeproceeding 0 0 0 Connect Cell Select Salinity mode by pressing the mode key until the correct...

Страница 8: ...repeated for each standard Return to the function you require by pressing the Mode key To Back off the background conductivity Follow points Oto 0 0 Place cell in background sample and enter 0 00 Pla...

Страница 9: ...r parameters for the same sam ple press the Mode key and then the Send key To print a new identifier press and hold down the Send key DATE OPERATOR SAMPLE COND lOOO S cm T 21 7 C 15 Computer Connect a...

Страница 10: ...Ifnot clean and repeat the process Wash with deionised water 0 Soak in deionised water for 48 hours before use Re check the cell constant before reusing 17 Section 11 Troubleshooting Symptom Probable...

Страница 11: ...adiustable 0 5 C 200mV 2 x 4mm sockets 9V Battery or power adaptor 210 x 150 x 88mm 550g 19 Section 13 Conductivity Cells Available from EDT Direct ION All cell listed have ATC Standard cable length i...

Страница 12: ...high conductivity e g Strong acid strong alkali strong salt solutions sea water Cell Conversion Table c s Units SI Units K I 0lcm K IO0 0lm K 0 llcm K I0lm K I0 0lcm K I000lm 21 Appendix 2 Calibration...

Страница 13: ...3 33 3 201 22 2 606 260 6 34 3 256 23 2 659 265 9 35 3 312 24 2 712 271 36 3 368 21 l 1 8mS cm 25 C 0 02M KC C mS cm mS m C mS cm 5 74 14 7 41 25 111 80 10 83 19 8 32 26 113 77 15 92 52 9 25 27 115 74...

Отзывы: