General  Information 

Read this section before proceeding with 

Sections 2,3,4,5 and 6 

Before calibrating, or making a measurement on any mode, it is 

important to note the 

+

The cell constant is correct

The temperature coefficient is correct

+

Any unnecessary calibration data is cleared.

+

Calibration data is indicated by the presence of the word

CAL on the display. To inspect the data, press the Chk key.

Calibration points are shown in sequence. To clear the data,

remain in CHK mode and press and hold down the Enter

key for 5 seconds.

+

The cell constant is not considered to be calibration data

but a physical constant. The value 

will not 

be erased by the

Clear procedure.

+

The temperature coefficient 

will 

be eased by the clear proce­

dure so always make a note of it and re-enter if necessary.

+

All measurements are automatically temperature compen­

sated to a base temperature of25

°

C.

+

To make un-compensated measurements then the tempera­

ture coefficient should be set to zero.

Section  2 

Conductivity 

Please refer to page 9 before proceeding. 

The following can only be carried out if the instrument has not 

been calibrated in Concentration Mode. 

Most conductivity cells are linear over a wide range. Since the cell 

constant has already been set (see p 7),  measure conductivity as 

follows: 

Connect cell and select conductivity mode 

El 

Place cell in sample 

Wait for a stable reading and record result noting the units. 

For very accurate results any non-linearity arising from polarisation 

effects, may be calibrated out as follows: 

Prepare up to four calibration solutions, each with a known 

conductivity 

El 

Connect cell and select conductivity mode 

Place cell in first standard and wait for a stable reading. 

Using the jJ, is keys, adjust the reading to the correct value 

(CAL flag will flash). Press Enter. (CAL flag will stop 

flashing). 

Repeat 0 for each standard. 

Place cell in sample and record value noting units. 

The calibration points may be inspected at any time as follows: 

CD 

© 

Press Clear 

First a number identifier is displayed and then the standard 

value. 

This is repeated for each standard. 

Return to the function you require by pressing the Mode 

key. 

10 

Содержание RE388TX

Страница 1: ...2 Salinity Section 6 12 Temperature Section 7 13 Concentration Section 8 14 The Recorder Output Section 9 15 Use with Printer or Computer Section 10 17 Platinizing Section 11 18 Troubleshooting Sectio...

Страница 2: ...have any ques tions about the shipment please call EDT Direct ION or your agent When you receive the shipment inspect the container for any signs ofdamage Note any evidence ofrough handling in transi...

Страница 3: ...cover J Battery Snap Figure 1 Bottom view ofinstrument showingpower connections 3 Instrument Test Procedure Refer to Figure 2 Switch the instrument on using the black switch on the back panel The mete...

Страница 4: ...vity l0MQ cm 0 MQ m Changing the multiple 0 Press the K key to display the multiple e Set the preferred units by pressing the K key to select either m orcm e Exit with the correct units set by pressin...

Страница 5: ...ESS THE Enter KEY WHEN SETTING THE CELL CONSTANT 7 Setting the Temperature Coefficient The conductivity ofelectrolytes changes with temperature The rate ofthis change is termed the temperature coeffic...

Страница 6: ...on 2 Conductivity Please refer topage 9 beforeproceeding The following can only be carried out ifthe instrument has not been calibrated in Concentration Mode Most conductivity cells are linear over a...

Страница 7: ...ding and record result noting units 11 Section 5 Salinity Salinity Measurement Please refer topage 9 beforeproceeding 0 0 0 Connect Cell Select Salinity mode by pressing the mode key until the correct...

Страница 8: ...repeated for each standard Return to the function you require by pressing the Mode key To Back off the background conductivity Follow points Oto 0 0 Place cell in background sample and enter 0 00 Pla...

Страница 9: ...r parameters for the same sam ple press the Mode key and then the Send key To print a new identifier press and hold down the Send key DATE OPERATOR SAMPLE COND lOOO S cm T 21 7 C 15 Computer Connect a...

Страница 10: ...Ifnot clean and repeat the process Wash with deionised water 0 Soak in deionised water for 48 hours before use Re check the cell constant before reusing 17 Section 11 Troubleshooting Symptom Probable...

Страница 11: ...adiustable 0 5 C 200mV 2 x 4mm sockets 9V Battery or power adaptor 210 x 150 x 88mm 550g 19 Section 13 Conductivity Cells Available from EDT Direct ION All cell listed have ATC Standard cable length i...

Страница 12: ...high conductivity e g Strong acid strong alkali strong salt solutions sea water Cell Conversion Table c s Units SI Units K I 0lcm K IO0 0lm K 0 llcm K I0lm K I0 0lcm K I000lm 21 Appendix 2 Calibration...

Страница 13: ...3 33 3 201 22 2 606 260 6 34 3 256 23 2 659 265 9 35 3 312 24 2 712 271 36 3 368 21 l 1 8mS cm 25 C 0 02M KC C mS cm mS m C mS cm 5 74 14 7 41 25 111 80 10 83 19 8 32 26 113 77 15 92 52 9 25 27 115 74...

Отзывы: