background image

9

Auto Range Digital Multimeter - USB Interface

3. F U N C T I O N   D E S C R I P T I O N S
3.1 General Functions 

HOLD mode

• HOLD mode makes the meter stop updating the display.
• Enabling the HOLD function in auto mode makes the meter switch to manual 

mode, but the full-scale range remains the same.

• Data Hold function can be cancelled by changing the measurement mode, 

pressing the R

RA

AN

NG

GE

E key, or push the H

HO

OL

LD

D/

/L

LIIG

GH

HT

T key again.

To enter and exit the ‘data hold’ mode: 
1. Press H

HO

OL

LD

D/

/L

LIIG

GH

HT

T key. (This fixes the display on the current value) the letter 

'H' will be displayed.

2. A second short press returns the meter to normal mode.

Manual and Auto range mode

The meter has both manual and auto (default) range options.
• In auto range mode, the meter selects the best range for the input detected. 

This allows you to switch test points without having to reset the range.

• In manual mode, you select the range. This allows you to override auto range 

and lock the meter in a specific range.

• When the meter is in the auto range mode, 'AUTO' is displayed.
To enter and exit the ‘manual’ range mode:
1. Press the R

RA

AN

NG

GE

E key. The meter will convert to manual mode. 'AUTO' will not 

show on the LCD. Each press of the RANGE key increments the range. When 
the highest range is reached, the meter wraps to the lowest range.

N

NO

OT

TE

E:: If you manually change the measurement range after entering the Data 

Hold modes, the meter exits this mode.

2. To exit the manual ranging mode, press and hold down RANGE key for 2 

seconds. The meter returns to auto mode and AUTO is displayed.

Battery Saver

The meter will engage a 'sleep mode' after 30 minutes if it is not used to save
battery life. Be aware data may be lost.
To come out of 'sleep mode':
1. Press the H

HO

OL

LD

D/

/L

LIIG

GH

HT

T key or rotate the rotary switch to wake the meter up. 

2. To disable 'sleep mode' hold down the SELECT key while turning the meter on.

5163440K_Instructions.qxd  18/09/2009  10:31  Page 9

Содержание DAM060

Страница 1: ...meter USB Interface OPERATING MANUAL Functions USB interface software for the PC Electrical testing Temperature testing from 5 to 1000 C ORDER CODE EDI 516 3440K MODEL DAM060 5163440K_Instructions qxd...

Страница 2: ...1 GENERAL INSTRUCTIONS 2 2 DESCRIPTION 6 3 FUNCTIONS DESCRIPTION 9 4 PC SOFTWARE OPERATION 16 5 TECHNICAL SPECIFICATION 17 6 MAINTENANCE 21 7 PACK CONTENT 22 5163440K_Instructions qxd 18 09 2009 10 3...

Страница 3: ...peration by measuring a known voltage and DO NOT use the meter if it operates abnormally Protection may be impaired When in doubt have the meter serviced Use the proper terminals function and range fo...

Страница 4: ...s when the meter displays you must change the battery Do not measure voltages above 600V in category III or 1000V in Category II installations When in REL mode the REL symbol is displayed Caution must...

Страница 5: ...detergent only ALWAYS set the dial to OFF position when meter is not in use If the meter is not going to be used for a long time take out the battery and do not store the meter in high temperature or...

Страница 6: ...6 2 DESCRIPTIONS 2 1 Instrument Familiarisation 1 LCD display 2 Keypad 3 Rotary switch 4 Terminals 5 USB adaptor port 1 2 5 3 4 5163440K_Instructions qxd 18 09 2009 10 31 Page 6...

Страница 7: ...er is in PC data transmission mode 6 Indicates Diode test mode 7 Indicates continuity check mode 8 Indicates Data Hold mode 9 R RE EL L Indicates Relative Measurement mode 10 M MA AX X Indicates maxim...

Страница 8: ...default to manual mode 2 Press RANGE to step through the ranges available for the selected function 3 Press and hold RANGE for 2 seconds 2 beeps will sound to return to auto ranging REL Press R RE EL...

Страница 9: ...nge In manual mode you select the range This allows you to override auto range and lock the meter in a specific range When the meter is in the auto range mode AUTO is displayed To enter and exit the m...

Страница 10: ...The meters voltage ranges are 600 0mV 6 000V 60 00V 600 0V and 1000V AC 600 0mV range only exists in manual mode To measure AC or DC voltage 1 Set dial to the D DC CV V A AC CV V or D DC Cm mV V range...

Страница 11: ...it is often different from the resistors rated value This is because the meters test current flows through all possible paths between probe tips In order to ensure the best accuracy in measurement of...

Страница 12: ...iode should still produce a forward bias reading of 0 5V to 0 8V however the reverse bias reading can vary depending on the resistance of other pathways between the probe tips Continuity Check To avoi...

Страница 13: ...e accuracy of measurements less than 60nF subtract the residual capacitance of the meter and leads Below 600pF the accuracy of measurements is unspecified Frequency and Duty Cycle measurement Do not m...

Страница 14: ...electrical shock do not use this instrument when voltages at the measurement surface exceed 60v DC or 24v rms AC To avoid damage or burns Do not make temperature measurement in microwave ovens To meas...

Страница 15: ...ds to the m mA A terminal for a maximum of 600mA For a maximum of 10A move the red test lead to the A terminal 5 Break the circuit path to be tested touch the black probe to the more negative side of...

Страница 16: ...lect As for the proper COM port we can view it in the device manager by following these steps Right click the My Computer icon on the Windows desktop and then click Properties Click the Hardware tab a...

Страница 17: ...LCD display Automatic indication of functions and symbols Range selection automatic and manual Over Range indication LCD will display OL Low battery indication The symbol is displayed when the battery...

Страница 18: ...1MHz Overload protection 250V DC or 250V AC rms RESISTANCE Overload protection 250V DC or 250V AC rms RANGE RESOLUTION ACCURACY 99 99Hz 0 01Hz 999 9Hz 0 1Hz 9 999Hz 0 001kHz 0 1 of rdg 3 digits 99 99...

Страница 19: ...ocouple errors Overload Protection 250V DC or 250V DC rms CAPACITANCE Overload Protection 250V DC or 250V DC rms RANGE RESOLUTION ACCURACY 55 C to 0 C 0 1 C 5 0 of rdg 4 C 1 C to 400 C 0 1 C 2 0 of rd...

Страница 20: ...es 10A DC or 10A DC rms for 10A ranges For measurements 6A 4 minutes maximum ON to measure 10 minutes OFF Above 10A unspecified RANGE RESOLUTION ACCURACY 600 A 0 1 A 1 5 of rdg 3 digits 6000 A 1 A 60m...

Страница 21: ...e the test leads and any input signals before opening the case Cleaning the multimeter Wipe the case with a soft damp not wet cloth and mild detergent DO NOT use abrasives or solvents Dirt or moisture...

Страница 22: ...r to unscrew the secured back cover This will expose the batteries 4 Replace the battery with a new 9V battery 1604 5 Replace the battery cover and screw securely back into place 6 The multimeter shou...

Страница 23: ...23 Auto Range Digital Multimeter USB Interface 5163440K_Instructions qxd 18 09 2009 10 32 Page 23...

Страница 24: ...developed and improved All details and illustrations are for guidance and may be subject to change EDISON The Power Factory PO Box 14 Wigston Leicester England ISO 9001 Registered Company 5163440K_Ins...

Отзывы:

Похожие инструкции для DAM060