User Guide

ELITE

Содержание HydraFacial MD Elite

Страница 1: ...2277 Redondo Ave Signal Hill CA 90755 USA Phone 800 603 4996 562 597 0102 Fax 562 597 0148 www edgeforlife com Made in the USA P N 19695 A Rev 4 18 2016 User Guide ELITE...

Страница 2: ...ents These intellectual property rights IPR s re main the exclusive property of Edge Systems or its licensors This product sold is subject to a license to use the IPRs only in conjunction with origina...

Страница 3: ...ctions 20 On Off Standby and Sleep Modes 21 Touchscreen 21 Adjust Fluid Flow 21 Adjust Vacuum Strength 22 Activate LED Lights optional 22 Skin Imaging Camera optional 23 System Cleaning Preventive Mai...

Страница 4: ...als clinics doctor s offices and in aesthetician s offices by trained doctors and or licensed aestheticians Please consult your state board for regulations The following is a tech nical description of...

Страница 5: ...Type BF Defibrillation Beweis BF Pi ce appliqu e de type D fibrillation preuve Direct Current Gleichstrom Courant continu Alternating Current Wechselstrom Courant alternatif ON Power EIN Stromversorgu...

Страница 6: ...ore susceptible to sunburn sun damage hyperpigmentation I should avoid excessive sun exposure especially between 10am 2pm _____ initial here 3 I have disclosed my history of allergies above and I ackn...

Страница 7: ...r Guide HYDRAFACIAL MD ELITE 7 SAFETY GUIDELINES 1 Ensure that all operators of the HydraFacial System are trained and licensed as required by the state or country Do not operate the unit before being...

Страница 8: ...USE MODIFICATION WHICH HAS BEEN USED IN VIOLATION OF THE SYSTEM INSTRUCTIONS OR FOR ANY PURPOSE OTHER THAN ONE FOR WHICH IT WAS MANUFACTURED OR CAUSED BY UNAUTHORIZED REPAIR OR USE OF UNAUTHORIZED PAR...

Страница 9: ...e HYDRAFACIAL MD ELITE 9 Verbraucher nderungen Unbefugte Verbraucher nderungen werden Garantie aufheben Edge Systems LLC wird verantwortlich f r irgendeine Verletzungen nicht sein die auf Grund unbefu...

Страница 10: ...inneren Teilen ist nicht notwendig Vermeiden Sie den Ge brauch um Agenten oder Chemikalien auf dem Instrument zu reinigen Einige Chemikalien d rfen Kunststoffteile oder Beschriftung besch di gen Irge...

Страница 11: ...ach service Clogging Flush all stations thoroughly Make sure the machine is cleaned after each service Gray cartridge is loose Make sure that the gray cartridge is properly inserted into the hand piec...

Страница 12: ...use Solution No Power Power cord Make sure power cord is properly connected to back of the unit and to a working wall outlet Fuses Unplug unit from main wall out let prior to this Use a small flat hea...

Страница 13: ...AGING CAMERA Clean the chin rest with an alcohol wipe after each use GETTING STARTED User Guide HYDRAFACIAL MD ELITE 13 Package Contents Console Unit Wet HydraFacial Handpiece Assembly Tool Holder Hyd...

Страница 14: ...LITE 27 Options WET DIAMOND AGGRESSIVE ABRASION MEDICAL USE ONLY 1 It is always recommended to use gloves for this process 2 After each use remove abrasive pads and tip Drop in hot water and let them...

Страница 15: ...e a plastic cleaner available at your nearest hardware store for example Novus plastic clean and shine Never under any circumstances use any type of alcohol to clean front door DGETTING STARTED User G...

Страница 16: ...until the arrow aligns with the lock symbol Note If the gray cartridge is difficult to insert drizzle a small amount of water into the cartridge to lubricate it and try again SYSTEM CLEANING User Guid...

Страница 17: ...onto the face of the hand piece held in an upright posi tion allowing the disinfectant to run into the handpiece opening before inserting the cleaning brush Note 70 isopropyl alcohol is not suitable...

Страница 18: ...in Solution Stations And Tubing AFTER EACH SERVICE 1 Wearing gloves is recommended 2 After each procedure run the system cleaning process Start by switching all the blue toggle switches down to the OF...

Страница 19: ...gainst the chin There are three options for capturing a photo In the member profile screen select the Photo Capture button or On the camera press the red button or Use the foot switch to capture a pho...

Страница 20: ...by mode TURN OFF THE SYSTEM 1 Follow the above steps to switch the machine into Stand by Sleep Mode Then switch the power entry module switch on the back of the unit to the OFF position and the machin...