Connector Signal Description 2/3
4. ENCODER CONNECTOR
Signal Name
Description
User matching signal
Synchro1
Synchro1 Input/Output
TTL 5V
-
If configured as OUTPUT (see in Capture software as below,
Type ≠ “1-axis eincrement”)
-
TTL 5V
-
= PRF = 5V/100ns pulse generated with each Pulser signal (IPEX
connector).
-
Signal can be used to synchronize an external system
-
Jitter = 10ns.
-
Absolute Max output c/-50 mA
-
If configured as INPUT (see in Capture software as below,
Type = “1-axis eincrement”)
-
TTL 5V, active on High level
-
Max Input current 50µA
-
When connected to Phased Array Wheel probe, Synchro1 is used as
« Index Increment » (short press) and « Scan Encoder Reset » (long
press > 2sec).
100 KΩ