LYNCS-CM Operation
1.
Remove one tractor assembly.
2.
Slide on any cable management guides required.
3.
Arrange tractor units and affix waterbox mount onto rail depending on required application
4.
Insert waterbox and secure into waterbox mount. The waterbox height will need to be adjusted
when on test item.
5.
Insert encoder cable and attach strain relief to strain relief point.
6.
Attach cable management sleeve to strain relief points.
7.
Attach LYNCS-CM Adapter lead to instrument and end of encoder cable.
8.
Setup data acquisition unit
–
for M2M instruments select LYNCS-CM
in scanner list NOTE: Typical
encoder resolution: 25.85 pts/mm.
9.
Perform encoder calibration and complete remaining setup configuration.
10.
Tether scanner from both handles.
11.
Place onto test item to perform scan.
12.
Start data acquisition on instrument and move scanner forward.
13.
Apply brakes and press button on tractor assembly to increment step on the instrument. Turn lever
lock to release pin from drilled hole and slide along rail, the pin will automatically drop into the next
hole.
14.
Release brakes and move scanner backwards. Repeat step 13 to build 2-Axis stitched C-Scan.
If not using an M2M instrument, the following changes in step 13 will need to be applied;
Step 13 - Apply brakes, increment step on the instrument, turn lever lock to release pin from drilled hole and
slide along rail, the pin will automatically drop into the next hole
NOTE: A long press on the tractor assembly button will reset the scan direction encoder to zero.