42
Inserting the IOM for Inhalable size fraction.
1.
Select Inhalable from the Aerosol library. See page 39 for further details.
2.
The SKC IOM is factory installed on the sensor. Simply remove any sampling inlets, if attached.
3.
Attach either a 25mm filter cassette or a 37mm filter cassette as shown below. Note: the orientation
of the sensor will need to be changed for 25mm sampling. A video on how to change the sensor
orientation is available on our website at
https://environmentaldevices.com/video-instruction/
Figure 3-5. Sampling for Inhalable Dust using 37mm or 25mm filter cassette.
Note:
If collecting concurrent 37mm filter samples place a gravimetric filter in the filter cassette. Pre-
weighed and preloaded filters cassettes of choice can also be used.
Note: For full instructions of how to change orientation see next page.
Inhalable Dust Particulates
Tubing
Attaches
Here
Position of
SKC IOM
with Adapter
Ring
Black knurled
thumbscrews
in this position