Using the Stimulator command (Setup menu) it is possible to send and
monitor pulses with a precision better than 1 ms,
. Pulses
up to about 30 s may be created by this method. For further details on
using the Stimulator in Chart refer to the
Chart Software Manual
.
Analysis of Chronoamperometry
Chronoamperomograms obtained from a planar disk electrode should
exhibit a current/time relationship governed by the Cottrell equation,
(see Appendix D for definition of terms)
It is usually easier to first simplify the Cottrell equation thus:
where
a
is treated as an empirical constant to be determined.
Stimulator settings to
produce a 500 mV high,
0.5 Hz square wave
Potential waveform generated by
these settings
i
nFAC D
1000
π
t
------------------------
=
i
a
t
-----
=
Figure 5–10
Simulator controls for
multiple step
chronoamperometry
(Macintosh).