SE-1200 Series Electrocardiograph Service Manual Inspection and Troubleshooting
- 32 -
Open the recorder casing and remove the recorder paper, and then the hint information
No Paper
should be displayed. Install the recorder paper again and close the recorder
casing, and then the hint information
No Paper
should disappear.
5.2 System Tests
The
System Test
screen is password protected and only technicians authorized by the
manufacturer can open it. The password is 006363.
1.
Opening the System Test Screen
When the
System Setup
screen is displayed, press
F2
on the keyboard to display the
Service
Password
window (Figure 5-1). After you enter the correct password and press
Enter
, the
System Test
screen (Figure 5-2) appears.
On the
System Test
screen, press
Tab
or
Shift + Tab
, or the Up, Down, Left or Right arrow to
move the cursor to a certain sub-item, and then press
Enter
to open the test screen related to the
sub-item.
Figure 5-1 Service Password Window
Figure 5-2 System Test Screen
Sub-item