SE-1200 Series Electrocardiograph Service Manual Optional and Advanced Functions
- 29 -
a)
Set
Auto Transmission
to
On
.
b)
FTP Information: Set
FTP User Name
and
FTP Password
to the user name and the
password you input in the
New User
window (refer to step 12 in configuring the
PC). Set
FTP path
to the subdirectory of the path you input in the
Default FTP Site
Properties
-
Home Directory
window (refer to step 8 in configuring the PC).
c)
IP Addresses: Set
Server IP
,
Local IP
,
Gateway
and
Subnet Mask
3)
Select a file format from the
File format
text box in the
File Setu
p window.
4)
Press
Enter
to confirm.
3.
Connect the electrocardiograph to the network of the PC with an Ethernet cable
recommended by the manufacturer.
4.
The electrocardiograph samples and analyzes the ECG data, and then transmits the ECG data
automatically to the folder you input in the
Transmission Setup
window.