CADENCE II Fetal Monitor User Manual
- 56 -
Appendix 1 Product Specifications
A1.1 Environmental Specifications
Temperature:
+5 ºC ~+ 40 ºC ( +41 ºF ~ +104 ºF)
Relative Humidity:
25% ~ 80% (non-condensing)
Working
Atmospheric Pressure:
860hPa ~ 1060hPa
Temperature:
-20 ºC ~ +55 ºC (-4ºF ~ +131 ºF)
Relative Humidity:
25% ~ 93% (non-condensing)
Transport and Storage
Atmospheric Pressure:
700hPa ~ 1060hPa
A1.2 Physical Specifications
Monitor
Size mm (depth x width x
height) :
330mm x 270mm x 100mm
Dimensions and Weight
Weight:
Approx. 3.6 kg
Operating Voltage:
100V-240V~
Operating Frequency:
50Hz/60Hz
Input Power :
60VA
Power Supply
Battery: 14.8V/6600mAh
Standards Compliance
IEC 60601-1:1988+A1+A2, EN 60601-1:1990+A1+A2,
IEC/EN 61157, IEC/EN 60601-2-37,
IEC/EN 60601-1-2:2001+A1
Anti-electric Shock Type
Class I equipment with internal power supply
Anti-electric Shock Degree
FHR1, FHR2, TOCO, FM, FS B
IUP BF
DECG CF
Degree of Protection against Harmful
Ingress of Water
Ordinary equipment (sealed equipment without liquid
proof)
Degree of Safety in Presence of
Flammable Gases
Equipment not suitable for use in presence of
flammable gases
Disinfection/Sterilizing Method
Refer to this user manual for details
EMC
Group I Class A