CADENCE II Fetal Monitor User Manual
- 25 -
5.2 Printing Configuration
CAUTION
:
Set the recorder well before printing starts. You can not change its configuration in the
process of printing.
5.2.1 Switching Auto-Print On and Off
1 Press the
SETUP
key six times to open the
Auto Monitor
menu.
2 Press the
CHANNEL
key when the cursor stays at
.
3 Press the
UP
key or
DOWN
key to toggle between
ON
and
OFF
(default).
4 Press the
CHANNEL
key to confirm selection.
5 Press the
AUTO
key to exit.
5.2.2 Choosing the Paper Speed
1 Press the
SETUP
key four times to open the
Print Configuration
menu.
2 Press the
CHANNEL
key when the cursor stays at
PSpeed
.
3 Press the
UP
key or
DOWN
key to choose the paper speed from 1cm/min, 2cm/min
(default) and 3cm/min.
4 Press the
CHANNEL
key to confirm selection.
5 Press the
AUTO
key to exit.
5.2.3 Choosing the Time Length
1 Press the
SETUP
key four times to open the
Print Configuration
menu.
2 Press the
UP
key or
DOWN
key to move the cursor to
Print TimeLen
; press the
CHANNEL
key.
3 Press the
UP
key or
DOWN
key to choose the print time length from 0 ~ 250 minute(s).
The default is 0. 0 means there is no time limit, the recorder will not stop until the
key is pressed.
4 Press the
CHANNEL
key to confirm selection.
5 Press the
AUTO
key to exit.
5.2.4 Switching Print Test On and Off
1 Press the
SETUP
key four times to open the
Print Configuration
menu.
2 Press the
UP
key or
DOWN
key to move the cursor to
PrintTest
; press the
CHANNEL
key.
3 Press the
UP
key or
DOWN
key to toggle between
ON
and
OFF
(default).
4 Press the
CHANNEL
key to confirm selection.
5 Press the
AUTO
key to exit.