WarmAir
Model 135 Manual
Cincinnati Sub-Zero Products, LLC
Page 4 of 17
Symbols
Voltage, Alternating Current
Potential Equalization Connection (Grounding)
Protective Earth (Ground)
Type BF Equipment (per IEC-601-1)
(Patient Applied)
Consult instructions for use and/or manual before operating
Over-Temperature Safety Limit
Under-Temperature Safety Limit
Hour Meter
Power On/Off Indicator
Fan Only Selection
Temperature Selection Switches
Low setting: 32.2°C
Medium setting: 37.8°C
High setting: 43.3°C
Separate collection for electrical and Electronic equipment. Dispose of the
WarmAir unit per hospital protocol.
CAUTION! Hose Nozzle MUST be connected to a compatible Forced Air
Blanket or thermal injury may occur.
WARNING: To prevent tipping when mounting the Model 135 unit to an IV
pole, clamp the unit no higher than 44 inches (112 cm) on an IV pole with a
minimum 24 inch (61 cm) diameter base. Failure to heed these restrictions
may result in IV pole instability, catheter site trauma, and patient/user
injury. Note: For instructions on use of the symbols, refer to the “Operating
Instructions” section
Insert Hose
<44
”
<112cm
>24
” >61cm