32
- Press the
WIND/+
and
PRESSURE/-
button to adjust values.
- Press the
SET
button to confirm & move to the next setting.
- Press the
ALARM
button to reset any adjusted value.
- Press the
LIGHT /SNOOZE
button at any time to exit.
6.9.2.1 Calibration Order:
1) Indoor temperature offset calibrated (range +/-5˚C, default: 0 degrees)
2) Indoor humidity offset calibrated (range +/-10%)
3) Outdoor temperature offset calibrated (range +/-5˚C, default: 0 degrees)
4) Outdoor humidity offset calibrated (range +/-10%)
5) Absolute pressure offset calibrated
(
± 10 hPa (± 2.95 inHg))
6) Wind direction offset calibrated (± 180 °)
7) Wind speed factor adjust, default 1 (range 0.5 to 1.5)
8) Rain factor adjust, default 1 (range 0.5 to 1.5)
6.9.3 Calibration Discussion
The purpose of calibration is to fine tune or correct for any sensor error
associated with the devices margin of error. Errors can occur due to
electronic variation (example, the temperature sensor is a resistive thermal
device or RTD, the humidity sensor is a capacitance device), mechanical
variation, or degradation (wearing of moving parts, contamination of
sensors).
Calibration is only useful if you have a known calibrated source you can
compare it against, and is optional. This section discusses practices,
procedures and sources for sensor calibration to reduce manufacturing and
degradation errors. Do not compare your readings obtained from sources
such as the internet, radio, television or newspapers. The purpose of your
weather station is to measure conditions of your surroundings, which vary
significantly from location to location.