104
5.2.7.4.6 Perform operations 5.2.7.4.5 for DER
)
10
(
0
p
H
=(900±100) mSv/h.
5.2.7.4.7 Maximum value of all received errors is
the limit of main relative permissible error of DER
measurement.
5.2.7.4.8 If the main relative permissible error limit
of DER measurement at 0.95 confidence probability
does not exceed:
- 20 % in the DER range of 1.0 to 10 μSv/h;
- 15 % in the DER range of 10 μSv/h to 1.0 Sv/h,