ECOTEST BDBG-09 Скачать руководство пользователя страница 35

Operating manual ВІСТ.418266.006-03 НЕ 

 

35 

 

“DER1 query”

 frame format –the data display system to the detecting unit  

D7  D6  D5  D4  D3  D2  D1  D0   

1  Byte 55h - start-of-frame character 

Byte AAh 

D7…D4–

protocol v1.3 character 

address 

D7…D0–

 detecting unit address * 

D7…D0–

“DER

query”

 frame code 

control 

arithmetical checksum with a carry 

 
* - 0FFh address – broadcast address. All detecting units respond to the query with this address. 

 

“Current DER1”

 frame format –the detecting unit to the data display system  

D7  D6  D5  D4  D3  D2  D1  D0   

1  Byte 55h - start-of-frame character 

Byte AAh 

D7…D4–

 protocol v1.3 character 

address 

D7…D0–

 detecting unit address 

D7…D0–

“Current DER1”

 frame code 

DER0 (low byte) 

DER, 

fixed point number, 

Least significant bit = 0.01 

µSv/

DER1 
DER2 

DER3 (high byte) 

Byte 

Statistical error of measurement 

D7 

D2  D1  D0 

D0,D1 - self-testing results of the 

detecting unit 

D0=1 - failure of the high sensitivity 

detector 

D1=1 - failure of the low sensitivity 

detector 

Reliable measurement result character 

D2=0 

 result is reliable 

D2=1 

 result is not reliable *

 

 

D7=0 

 LSB DER = 0,01 

µSv/h

 

D7=1 

 LSB DER = 0,1 

µSv/h

 

control 

arithmetical checksum with a carry 

 

* - measurement result is accepted as not reliable if the statistical error of measurement exceeds maximum permissible 
error of measurement 

Содержание BDBG-09

Страница 1: ...BDBG 09 DETECTING UNIT OF GAMMA RADIATION Operating manual...

Страница 2: ...Operating manual 418266 006 03...

Страница 3: ...3 3 CONTENTS 1 DESCRIPTION AND OPERATION 4 2 PROPER USE 10 3 MAINTENANCE 13 4 REPAIR 20 5 STORAGE AND PUTTING IN PROLONGED STORAGE 20 6 SHIPPING 20 7 DISPOSAL 20 APPENDIX 21 APPENDIX B 28 APPENDIX C 4...

Страница 4: ...Technical specifications 1 2 1 Key specifications are presented in the Table 1 1 Table 1 1 Key specifications of the detecting unit Name Unit of measurement Standardized value according to the specifi...

Страница 5: ...1 Concerning the resistance to climatic and other environmental factors the detecting unit meets the requirements outlined below 1 2 2 2 The detecting unit is resistant to the influence of the followi...

Страница 6: ...kraine No 175 dated October 17 2014 design elevation up to 70 m 1 3 Delivery kit of the detecting unit The delivery kit of the detecting unit consists of units and maintenance documentation given belo...

Страница 7: ...nt 3 and corbel 4 to attach to the wall Figure 1 a Alternatively the detecting unit can be fastened to the surface where it will be directly installed is shown in Figure 1 b The Figure 1 b shows that...

Страница 8: ...ow with the account of personal background of the detectors The microcontroller also calculates a maximum value of DER measurement statistical error for each gamma radiation DER value At the same time...

Страница 9: ...001 468353 001 Serial port adapter 436231 001 Power unit 685621 002 Service cable 685622 002 Service cable 467371 001 CD with technological software 7 Personal computer IBM compatible personal comput...

Страница 10: ...d into unitized shipping containers boxes Inside surfaces of the walls bottom and cover of the box should be furnished with corrugated cradboard Note Other types of unitized containers containers cart...

Страница 11: ...nnect the detecting unit to the adapter with the help of the 685621 002 service cable 2 2 3 3 Switch the PC on connect the power unit of the adapter to 220 22 V industrial network Start the bdbg exe p...

Страница 12: ...during use of the detecting unit fully comply with the requirements presented in 2 2 1 of the OM 2 3 1 2 The direct use of the detecting unit is not dangerous for the maintenance personnel and is env...

Страница 13: ...ng unit 3 1 3 1 External examination 3 1 3 1 1 External examination of the detecting unit should be performed in the following order check the technical condition of the detecting unit surface integri...

Страница 14: ...uate the necessity of sending the detecting unit for repair and type of repair for mid life repair deviation of parameters from control values during periodical verification of the detecting unit b mi...

Страница 15: ...e 467371 001 CD with technological software Usage of other measurement equipment that meets the specified accuracy is allowed 3 2 3 Testing conditions Testing should be carried in compliance with the...

Страница 16: ...R by the formula 5 10 5 1 i i H H 3 1 3 2 4 3 6 Place the 3 carriage with the detecting unit in a position where gamma radiation DER from 137 s source is equal to 10 0 H 0 8 0 1 Sv h 3 2 4 3 7 In thre...

Страница 17: ...hat measurement results should be taken after the probable value of measured gamma radiation DER appears in the MEASURED RATE zone of the bdbg exe program window Value probability is indicated by gamm...

Страница 18: ...e and with 6 0 0 1 V voltage connect to the terminals 12 V and 6 V of the switching desk respectively 3 2 4 3 22 2 Fix the detecting unit in the 3 carriage holder so that the mechanical center of gamm...

Страница 19: ...3 22 13 Perform operations 3 2 4 3 22 6 3 2 4 3 22 7 and 3 2 4 3 22 9 for gamma radiation DER from 137 Cs source 137 Cs 10 0 H 800 0 100 0 Sv h 3 2 4 3 22 14 Perform operations 3 2 4 3 22 6 3 2 4 3 2...

Страница 20: ...olonged storage or repeated use silicagel should be dried Total time of the detecting unit storage with the account of the repeated closing down should not exceed 10 years 6 SHIPPING 6 1 The detecting...

Страница 21: ...logy software includes bdbg exe program drivers for USB serial device and serial port 6 Installing technology software Installation requires administrator rights 6 1 Installing drivers for the USB ser...

Страница 22: ...418266 006 03 22 For proper functioning of the bdbg exe program set the minimum value of 1mS in the property of Latency Timer of the USB serial port Do the following open the Properties window USB Se...

Страница 23: ...Operating manual 418266 006 03 23 select the Port Settings tab click Advanced change the value of Latency Timer ms complete the input by clicking...

Страница 24: ...an be calculated using the Windows command line certutil hashfile bootdrive Program Files SVC bdbg exe MD5 If the calculated MD5 hash does not match with the above one the BDBG EXE program has been co...

Страница 25: ...on the monitor Serial indicates a serial number of the detecting unit MEASURED DOSE RATE indicates DER measurement result received from the detecting unit It can be indicated in two colors grey and b...

Страница 26: ...nge If the detecting unit is operated using extended protocol v1 3 the current response delay factor to the broadcast query is displayed in the Current delay factor field This factor is also stored in...

Страница 27: ...g the detecting unit receiving and displaying DER measurement results Queries are sent to the detecting unit with one second interval Stop clicking this button terminates the exchange process started...

Страница 28: ...ng unit will respond with the Factors frame where the current calibration factors will be given To change the calibration factors the data display system should transmit the New factors frame to the d...

Страница 29: ...Byte AAh 1 0 1 0 1 0 1 0 0 0 0 1 3 2 1 0 D7 D4 Current DER frame code D3 D0 detecting unit address DER0 low byte DER fixed point number Least significant bit LSB 0 01 Sv h DER1 DER2 DER3 high byte Byt...

Страница 30: ...st address All detecting units respond to the query with such address Current temperature frame format the detecting unit to the data display system D7 D6 D5 D4 D3 D2 D1 D0 0 1 0 1 0 1 0 1 Byte 55h st...

Страница 31: ...sa middle byte mantissa high byte Personal background of the high sensitivity detector float MSP430 exponent mantissa low byte mantissa middle byte mantissa high byte Multiplication factor of the low...

Страница 32: ...e low byte Dead time physical of the high sensitivity detector word high byte low byte Dead time physical 3 of the high sensitivity detector word high byte mantissa high byte Background K of the high...

Страница 33: ...ial No _0 low byte Serial No of the detecting unit Serial No _1 Serial No _2 Serial No _3 high byte control arithmetical checksum with a carry Address change frame format the data display system to th...

Страница 34: ...the response delay factor to broadcast query the data display system should send the Serial _1 query frame to the detecting unit The detecting unit will respond in 5 ms to 15 ms with the Serial _1 fra...

Страница 35: ...start of frame character Byte AAh 1 0 1 0 1 0 1 0 0 1 1 1 0 0 0 0 D7 D4 protocol v1 3 character address D7 D0 detecting unit address 0 0 0 0 0 0 0 1 D7 D0 Current DER1 frame code DER0 low byte DER fix...

Страница 36: ...FFh address broadcast address All detecting units respond to the query with this address Current temperature1 frame format the detecting unit to the data display system D7 D6 D5 D4 D3 D2 D1 D0 0 1 0 1...

Страница 37: ...hysical 3 of the high sensitivity detector word high byte mantissa high byte Background K of the high sensitivity detector float MSP430 exponent mantissa low byte mantissa middle byte mantissa high by...

Страница 38: ...1 0 0 0 0 0 0 1 1 D6 D0 Confirmation1 frame code D7 0 normal operation D7 1 error control arithmetical checksum with a carry Current factors1 query frame format the data display system to the detecti...

Страница 39: ...l 3 of the high sensitivity detector word high byte mantissa high byte Background K of the high sensitivity detector float MSP430 exponent mantissa low byte mantissa middle byte mantissa high byte Per...

Страница 40: ...ess broadcast address All detecting units respond to the query with such address Serial _1 frame format the detecting unit to the data display system D7 D6 D5 D4 D3 D2 D1 D0 0 1 0 1 0 1 0 1 Byte 55h s...

Страница 41: ...code new address D7 D0 new address of the detecting unit new constant D7 D0 new response delay factor to broadcast query control arithmetical checksum with a carry Confirmation1 frame format the detec...

Страница 42: ...and v1 3 protocol is calculated according to Figure B 1 bytei D7 D6 D5 D4 D3 D2 D1 D0 arithmetical summation checksumi 1 D7 D6 D5 D4 D3 D2 D1 D0 partial result carry D7 D6 D5 D4 D3 D2 D1 D0 arithmeti...

Страница 43: ...IRSCHMANN interface connection serves to connect the data display system to the detecting unit The interface connection contains the following signals Signal contact RS 485 Circuit A 1 RS 485 Circuit...

Страница 44: ...d time formula 50 Dead time physical 49 642 BDBG 09 0308123 Response delay factor to broadcast query 1 06 04 2009 15 14 34 0 00 Sv h 255 OK 06 04 2009 15 14 44 0 12 Sv h 115 OK 24 3 06 04 2009 15 14 5...

Страница 45: ...n CA6LD connection resistance to mechanical and climatic factors depending on the use conditions active resistance per unit length depending on the cable length according to the formula D 1 it is dete...

Страница 46: ...plane 100 80 60 40 20 0 Cs 137 60 241 1 90 0 150 0 180 0 0 0 270 0 30 0 120 0 60 0 3300 3000 240 0 210 0 241 Cs 137 Co 60 Anisotropy of the BDBG 09 horizontal plane 100 80 60 40 20 0 Cs 137 60 241 1 9...

Отзывы: