Installation/Decommission
Page 49
Diagnostics Menu
Press/Temp/Flow Comp:
ON
Diagnostic mode:
Operate
Control Loop:
Enabled
2.6
Transporting/Storage
Transporting the instrument should be done with great care. It is recommended that the packaging
the instrument was delivered in should be used when transporting or storing.
When transporting or storing the instrument the following points should be followed:
1.
Turn OFF the instrument and allow it to cool down.
2.
Remove all connections of pneumatic pipe, power cables and communication cables from the
instrument.
3.
Refer to Figure 38. If storing over a long period (six months) turn the battery OFF by moving the
switch (S1) on the main controller PCA.
4.
Remove the instrument from the rack.
5.
Refer to Figure 39. Seal each pneumatic ports with a dust plug.
Figure 39
–
Dust Plugs
6.
Refer to Figure 37. If transporting the instrument, replace the travel screws.
CAUTION
Major damage to the instrument components may occur in transit if the
measurement cell is not secured in place with the supplied transit screws.
7.
Refer to Figure 36. Remove the USB memory stick and pack with instrument.
8.
Place the instrument back into a plastic bag with desiccant packs and seal the bag (ideally the bag
supplied upon delivery).
Содержание Acoem Serinus 30
Страница 2: ...Serinus 30 User Manual 3 3 Page 2 This page is intentionally blank...
Страница 20: ...Serinus 30 User Manual 3 3 Page 20 This page is intentionally blank...
Страница 38: ...Serinus 30 User Manual 3 3 Page 38 Figure 22 Received Item...
Страница 118: ...Serinus 30 User Manual 3 3 Page 118 This page is intentionally blank...
Страница 138: ...Serinus 30 User Manual 3 3 Page 138 This page is intentionally blank...
Страница 144: ...Serinus 30 User Manual 3 3 Page 144 Figure 131 Avoid to Touching Particulate Filter Paper...
Страница 157: ...Serinus 30 User Manual 3 3 Page 157 This page is intentionally blank...
Страница 162: ...Serinus 30 User Manual 3 3 Page 162 7 1 Flow Fault Figure 142 Flow Fault Diagnostic Procedure Section 2 3 2 Section 5 2...
Страница 165: ...Troubleshooting Page 165 7 4 Mirror Temperature Failure Figure 145 Mirror Temperature Failure Diagnostic Procedure...
Страница 166: ...Serinus 30 User Manual 3 3 Page 166 7 5 USB Memory Stick Failure Figure 146 USB Memory Stick Failure...
Страница 169: ...Serinus 30 User Manual 3 3 Page 169 This page is intentionally blank...
Страница 192: ...Serinus 30 User Manual 3 3 Page 192 9 5 Plumbing Schematic PN D020005...
Страница 193: ...Parts List and Schematics Page 193 9 6 Block Wiring Schematic PN D020102...
Страница 194: ...Serinus 30 User Manual 3 3 Page 194 9 7 Measurement Cell PN H014100...
Страница 195: ...Parts List and Schematics Page 195 9 8 Source Timing Assembly PN H014125...
Страница 196: ...Serinus 30 User Manual 3 3 Page 196 9 9 Calibration Manifold Assembly PN H010013 01...
Страница 197: ...Parts List and Schematics Page 197 9 10 Valve Assembly PN H010042...
Страница 198: ...Serinus 30 User Manual 3 3 Page 198 9 11 High Pressure Valve Exploded View PN H050043...
Страница 199: ...Parts List and Schematics Page 199 9 12 Sample Filter Holder Assembly PN H010160...
Страница 200: ...Serinus 30 User Manual 3 3 Page 200 9 13 CO CO2 Converter Assembly PN H014130...
Страница 201: ...Serinus 30 User Manual 3 3 Page 201 This page is intentionally blank...