Range 1
– AMS100
Operating instructions
ECOSAFE | Date du document : 29/04/2019 – Version : 01
3
▪
That it is protected against moisture and that the ambient temperature of the location is neither less than 5°C, nor more than
40°C
Transport and handling
The safety cabinet that you received has been protected and transported in optimal conditions, as to be protected from any damage.
It has to be lifted only by suitable equipment (suct
ion cup, strap…). It has to be carried in a vertical position, and it must not be lying.
It is imperative not to tilt the cabinet when it is carried. Indeed it can cause damages that can alter the right functioning of the cabinet.
On the other hand, if it has to be moved, the cabinet has to be placed on a pallet and strapped, as to avoid any fall.
Commissioning
After having unwrapped the cabinet, ensure that you also received a set of two keys.
Your safety cabinet has to be installed, used and maintained as to guarantee the users protection.
4. INTERNAL EQUIPMENTS
The cabinet is provided with 3 perforated shelves. These must be fixed at the chosen height thanks to the rack and pinion. Take care
not to exceed the loading capacity per shelf of 50 KG.
The cabinet also has an integrated retention tank, provided with a drain plug positioned under the cabinet.
It should not be used as a storage area. Its retention capacity is 112 liters.
5. STORAGE
Your safety cabinet is exclusively destined to store chemicals in laboratories, industries, schools
… in closed containers.
It must not store high concentrated corrosive products.
We remind you that the storage of corrosives can alter the quality of your safety equipment. On the other hand, we remind you that
damages due to a bad use of the equipment and in particular to corrosion won
’t be covered by our warranty.
6. VENTILATION
We highly recommend connecting your safety cabinet to a ventilation system, as to evacuate the eventual vapors contained in it.
The air renewal has to be at least 10 times the volume of the cabinet, per hour. Moreover, ensure that the ventilation runs 24h/7.
Connection to a ventilation system
The air inlet and outlet are situated on the side of the cabinet. See the below outline.