DIAGNOSTIC X-RAY UNIT 

         

               

User Manual 

  ORANGE 1040HF 

    - 

ULTRA 1040HF, ULTRA 100HF

 

     

 

 

 

 

 

 

 

 

 

           
 
 

Document No. : EUM-Vet-01 

Issued Date : May.26,2016   

Revision : 1.1

 

 

 

  For better performance and safety, this manual may be changed without any prior notice. 

The original version of ORANGE 1040HF is being written in English. 

Содержание ORANGE 1040HF

Страница 1: ...1040HF ULTRA 1040HF ULTRA 100HF Document No EUM Vet 01 Issued Date May 26 2016 Revision 1 1 For better performance and safety this manual may be changed without any prior notice The original version of ORANGE 1040HF is being written in English ...

Страница 2: ...bols and terms Product name designation 13 13 13 14 Electromagnetic compatibility statement 14 Cleaning 14 Proper environment 15 Accessories composition 16 Dimension 17 Name of each part Identification label Operation panel Specifications X ray tube specification 18 19 20 21 22 Section 2 Operation Operation conditions 25 Operation instruction 26 Error code and corrective actions Exposure data tabl...

Страница 3: ...output Lamp exchange procedure Fuse exchange procedure Put in the hand switch Put in the power connector and plug Diagnostic x ray unit stand fixing 34 35 38 39 40 40 41 Section 4 Schematics Block diagram Power board Power board DRV Main board CPU Main board EP_FB Main board IP_FB 44 45 46 47 48 49 Table of Contents ...

Страница 4: ... 4 Diagnostic X ray Unit USER MANUAL Section 1 General Information ...

Страница 5: ... A1 2010 EMC EN 60601 1 2006 SAFETY EN 60601 1 3 2008 SAFETY Diagnostic X ray Unit with radiation protection in accordance with EN 60601 1 3 2008 EN 60601 2 28 2010 SAFETY X ray Tube Assembly ORANGE 1040HF HV Tank EN 60601 2 28 2010 EN 60601 2 54 2009 SAFETY X ray Equipment for Radiography ORANGE 1040HF EN 60601 2 54 2009 Risk Management Report Report Number GRM 01 Complied Standard ...

Страница 6: ...g gu Seoul Korea TEL 82 70 7510 3400 FAX 82 70 8630 3420 E mail sales ecoray kr sales ecoviewmedical com URL www ecoray kr EC authorized representative BCF Technology Ltd Address Imaging House Phoenix Crescent Strathclyde Business Park Bellshill ML4 3NJ Scotland UK TEL 44 0 1506 460023 FAX 44 0 1506 460045 Contact ...

Страница 7: ...ide factors such as fire flood storm tidal wave lightening earthquake theft abnormal conditions of operation and intentional destruction of the system Insubstantial defects do not qualify refund e g prices for the batteries training materials and supplies are not covered We do not take responsibility for damages or loss which appear after the guarantee period The guarantee does not cover additiona...

Страница 8: ...pt those specified in this manual for operation If you have experienced any trouble with the equipment the main switch off immediately and contact its authorized dealer for the service for assistance Pregnant animals should be avoid radiation and consult with a doctor if required Do not use the other connector which it does not provided from manufacturer The equipment is designed for the purpose o...

Страница 9: ...n of radiation Otherwise the extra object can make unexpected effect to X ray image that may interfere correct diagnosis Eliminate all the useless objects from the zones because they are a further source of secondary radiation above all near the source The focal to skin distance shall be not shorter than 45 cm Operators are requested to use minimum kV and mAs values for experimental use Do not exc...

Страница 10: ... of both animals and radiologists therefore Must use the diagnostic x ray stand A way to fixed Page 41 To avoid risk of electric shock this equipment must only be connected to a supply mains with protective earth Do not modify this equipment without authorization of the manufacturer Check that the input power frequency voltage and allowable current matches with the unit requirements This unit has ...

Страница 11: ... of dangerous voltage This label placed on the top of HV Tank HV Tank uses high voltage to generate X ray so cut off the power before checking and repairing the unit Warning of ionized Radiation This label placed on the top of top case The ionized radiation is dangerous for the animal and the operator if the appropriate safety measures are not strictly observed Warning Label Information ...

Страница 12: ...on the top of HV tank This label explained HV tank information Protective Earth This label placed on the inside of case Identify the point where the system safety ground is fastened to the chassis Protective earth is connected to conductive parts of Class Ⅰ equipment for safety purposes Hand Switch Information This label placed on the side of Hand switch This label explained manufacturer and model...

Страница 13: ...the manufacturer or his agent from responsibility for any damage or injury as a result of the system noncompliance The responsibility about use of medical apparatus and management of maintenance is in a user side Symbols and Terms The following symbols will be used in this manual their meaning and their application are described further below Caution Calls attentions to action or conditions that c...

Страница 14: ...ce which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by on or more of the following measures 1 Reorient or relocate the receiving device 2 Increase the separation between the equipment 3 Connect the equipment into an outlet on a circuit different from that to which the other devices are connected If the problem continues consult t...

Страница 15: ...nsportation condition you must maintain the following range of temperature and humidity and atmosphere Storage Transport Environment Temperature range 25 C 60 C Relative Humidity Range 10 95 RH Air Pressure Range 500 1060 hPa non condensing not packaged for outdoor storage Caution Avoid the following environments for operation or storage Where the equipment is exposed to water vapor Where the equi...

Страница 16: ...ed one body X ray equipment therefore following regions are firmly fixed and connected of consist of Picture Name Specification Hand Switch Manufacture OMRON Model Name C2U Push button switch 2 Steps 1 Step Ready 2 Step Shot Extra Lamp Manufacturer OSRAM Model Name 64610HLX Power 12 V 50 W Extra Fuse Body Type Ceramic Power 15 A 250 V User Manual Size W150 X H120 mm Aluminum Carrying Case Shoulder...

Страница 17: ... 17 Diagnostic X ray Unit USER MANUAL Dimension Unit mm Without accessories Weight 12 5 kg ...

Страница 18: ...8 Diagnostic X ray Unit USER MANUAL Name of Each Part Handle Laser Beam Outlet Main Body Angle Meter Operation Panel Measuring Tape Collimator Main Switch Collimator Knob Power Inlet Hand Switch Mono Jack ...

Страница 19: ...art To identify a type B applied part complying with IEC 60601 1 Note B Body Disposal Disposal of this product must be handled according to local laws and regulations Date of manufacture Date of product manufacture Manufacturer Information Basic information of manufacturer AR Information Authorized Representative in European Community Caution General caution CE Mark Conformite European ...

Страница 20: ...is being made It It lamp on during 5 sec after exposure It lamp on occurring is system problem kV Key kV can be set mAs Key mAs can be set APR Key It is available for 16 8 APR data Memory kV mAs Display reversal key It can be kV mAs reverse display Collimator Lamp It indicates a field to be exposed Laser beam key For exposure focus Store key for APR Save the selected APR data ...

Страница 21: ... ray tube Inherent filtration Fixed added filter AL filter Acrylic window Collimator Total filtration External electrical power source Class I Type B Applied Part IPX 0 Ordinary equipment Continuous operation with intermittent Time interval between 3 exposure rest time 5 min Europe 220 240 V Americas 110 120 V Single Phase Europe 50 Hz Americas 60 Hz 3 8 kVA 80 kV 30 mA Less than 2 80 kV 30 mA Les...

Страница 22: ... mm Target Angle 16 degree Target Material Tungsten Inherent Filtration At least 0 8 mmAl at 50 kV Insulating Medium In Oil Nominal Voltage 40 100 kV Anode Heat Content 35 kJ 50 kHU Max Anode Heat Dissipation 250 W 350 HU s Nominal Anode Input Power at 1 0s See rating charts Weight About 430 g X ray Tube Outline Dimension ...

Страница 23: ... 23 Diagnostic X ray Unit USER MANUAL Rating Chart ...

Страница 24: ... 24 Diagnostic X ray Unit USER MANUAL Section 2 Operation ...

Страница 25: ...e operator kV mAs kV The operator establishes the value the system chooses automatically the maximum value of mA in order to obtain the maximum radiological result mAs The operator establishes the value the system chooses automatically the minimum value in order to the maximum radiological result Caution Connection to the electric energy before the electric connection make sure that the voltage an...

Страница 26: ...from the unit or to the full extension of the hand switch cord before making exposure 3 Always use the proper field size and technique factors for each procedure to minimize X ray exposure and produce the best diagnostic result 4 Check the digital display carefully before making an exposure verify that the selected technique is the intended technique Pay particular attention to the placement of th...

Страница 27: ...elow and check it exposes correctly Review Execution of Radiographs in page 29 for reference of procedure of x ray exposure every 3 exposure take the rest of 5 minutes The below chart also uses for x ray tube aging 6 Positioning the device at the ready condition by one time pushing of hand switch 7 When the READY LED Lamp Laser lighted press the hand switch again and make exposure During the x ray...

Страница 28: ...y at the same time 3 8 Mode Store To switch APR mode to 8 Mode when the main switch is OFF press and hold 1 and 4 key at the same time and then turn on the main switch In 8 Mode the user can set kV and mAs values in each numbers through 1 to 8 To save click the number from 1 to 8 and set kV and mAs value by using UP DOWN keys and press and hold the STORE key until you hear buzzer sounds 4 Mode Cha...

Страница 29: ...g an exposure press the hand switch half way down During 1 5 seconds after this first press the filament is heated and green lamp READY lights Now depress the hand switch fully down and the exposure is given immediately The duration of the exposure is signaled by the orange lamp X RAY and an audible X ray buzzer sound at the exposure Note 1 The exposure must be made pressing the exposure switch in...

Страница 30: ... is displayed when the Over Current Protect is detected Reset main switch If proper operation is unavailable replace MAIN BOARD or replace PWR BOARD Err06 It is displayed when the IP FB and EP FB signal is high in the state of exposure Reset main switch If proper operation is unavailable replace MAIN BOARD Err07 It is displayed when the mAs error Reset main switch If proper operation is unavailabl...

Страница 31: ...8 10 11 66 1 6 No grid 74 0 4 No grid 60 1 0 12 13 70 1 6 78 0 8 64 2 0 14 15 72 1 6 80 0 8 68 2 5 16 17 72 2 0 86 0 8 68 2 5 18 19 74 2 4 90 0 8 70 3 2 20 21 76 2 4 90 1 2 74 3 2 22 23 80 2 4 96 1 2 80 4 0 26 66cm Distance Film DIS G400 Cassette Screen Rare Earth Regular DIS G400 Caution This exposure data table can be set to different kV and mAs accordingly a animal body situation Therefore the ...

Страница 32: ...indicates any significant zone of occupancy designated in accordance with this subclause shall have a floor area of dimensions not less than 60 cm x 60 cm and a height of not less than 200 cm and wear appropriate lead apron to protect himself from leakage radiation h Height above floor 1 100kV with protective device 2 70kV with protective device 3 100kV without protective device 4 70kV without pro...

Страница 33: ... 33 Diagnostic X ray Unit USER MANUAL Section 3 Service Manual ...

Страница 34: ... TP6 VR3 11 If Channel 2 kV does not display correct value of 2 0 V turn VR3 on the main board Clockwise Increase Anti clockwise Decrease 12 In case mA Channel 1 Overshoot peak value is more than 8 2 V in the 8 V standard decrease it to meet at 8 5 V by using VR5 Clockwise Decrease Anti clockwise Increase 13 Check the voltage level on each kV step Undershoot can be found at high kV ranges for exam...

Страница 35: ...SER MANUAL Reproducibility of the Radiation Output 1 Accuracy of X ray tube voltage Caution When Calibration is required it must be done by X ray service technicians or manufacture The Calibration required period is every 2 years ...

Страница 36: ... 36 Diagnostic X ray Unit USER MANUAL 2 Accuracy of current time product ...

Страница 37: ... 37 Diagnostic X ray Unit USER MANUAL ...

Страница 38: ...ottom of a unit 3 Remove two bolts B of Lamp gobo 4 Remove Lamp gobo C 5 It checks that the inside temperature has fully got low 6 A glove is worn the lamp D is pulled horizontally and is removed 7 A new lamp is inserted in a socket 8 The gobo C and the plate bolts A for lamp exchange are attached in the order of the opposite procedure of 3 4 and 5 A B C D ...

Страница 39: ...e 1 It checks that the main switch turned off 2 Remove four bolts A at the Upper cover of a unit 3 Open the Upper cover 4 A new fuse B is inserted in a fuse socket 5 After inserted a new fuse in a fuse socket Cover Upper cover of a unit and joint four bolts A B ...

Страница 40: ...of action by pressing the button 1 Step press button Lamp and Laser ON 2 Step press button Shoot X ray Put in the Power Connector and Plug 1 Before put on the power cable connector B in power socket A Check a foreign substance in connector hole C 2 If put on the power cable connector B and plug in power socket A and point Don t use wet hands A A C B ...

Страница 41: ...re and make it stable by tighten the bolt 2 Before purchasing the stand check the size of captive bolt is 8 mm Note 1 When collimator lamp and fuse are needed to be replaced replace them by following the manual The each spare collimator lamp and fuse are stocked inside the aluminum carrying case 2 When x ray SID and dual laser pointer adjustment are needed contact manufacture for adjustment ...

Страница 42: ... 42 Diagnostic X ray Unit USER MANUAL Fixed Form Dimension ...

Страница 43: ... 43 Diagnostic X ray Unit USER MANUAL Section 4 Schematics ...

Страница 44: ... 44 Diagnostic X ray Unit USER MANUAL Block Diagram ...

Страница 45: ... 45 Diagnostic X ray Unit USER MANUAL Power Board ...

Страница 46: ... 46 Diagnostic X ray Unit USER MANUAL Power Board DRV ...

Страница 47: ... 47 Diagnostic X ray Unit USER MANUAL Main Board CPU ...

Страница 48: ... 48 Diagnostic X ray Unit USER MANUAL Main Board EP_FB ...

Страница 49: ... 49 Diagnostic X ray Unit USER MANUAL Main Board IP_FB ...