background image

Instruction Manual

Model No: ECO-NEB01

COMPLETE NEBULISER KIT

(ADULT & PEDIATRIC MASK WITH MOUTHPIECE)

NEBULISER

Compact
Lightweight
Family Friendly

NEBULISER

booklet_REV_042020.indd   3

2020/04/06   11:43

Содержание ECO-NEB01

Страница 1: ...Instruction Manual Model No ECO NEB01 COMPLETE NEBULISER KIT ADULT PEDIATRIC MASK WITH MOUTHPIECE NEBULISER Compact Lightweight Family Friendly NEBULISER oklet_REV_042020 indd 3 2020 04 06 11 4...

Страница 2: ...ations 3 Description 4 Accessories 4 Operation Instruction 5 Cleaning 5 Maintenance 6 Warranty 6 IEC Symbols 6 Important Safeguards 6 Technical data 8 EMC 9 Troubleshooting 13 Disposal 14 Manufacturer...

Страница 3: ...Hz Power Consumption Below 65W or 60W Rated Current Below 1 5A or 0 70A Dimensions Length 155mm Width 140mm Height 95mm Weight 1 2kg 2 64lbs Sound Pressure Level 58 dB A Max Comp Pressure 35 Psi to 50...

Страница 4: ...hpiece 8 Tubing 3 Power switch 6 Nebuliser cup 9 Foam filters Accessories Nebuliser machine Air tube Adult mask Paediatric mask Nebuliser cup T Mouthpiece 5pcs air filters Power switch Air Outlet Conn...

Страница 5: ...the top of the nebuliser cap 7 Connect tubing to nebuliser air inlet connector 8 Turn the power on to start the unit 9 Implement the treatment by placing the mouthpiece between the teeth Inhale and ex...

Страница 6: ...s cotton Do NOT operate without a filter Warranty We warrant the Econocare nebuliser to be free from defective workmanship and materials for a period of 1 year from the date of purchase Check with you...

Страница 7: ...off when it is not in use 6 Keep the cord away from HEATED or HOT surfaces 7 NEVER drop or insert any object into any opening 8 NEVER block the air openings of the product or place it on a soft surfa...

Страница 8: ...in gress of water is Ordinary Technical data Particle Size MMAD approximately 3 um Appropriate Medication Quantities 2ml minimum to 6ml maximum Nebulisation rate 0 33ml min by weight loss Aerosol Outp...

Страница 9: ...served to verify normal operation in the configuration in which it will be used Guidance and manufacture s declaration electromagnetic emission This is intended for use in the electromagnetic environm...

Страница 10: ...of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT f...

Страница 11: ...ss than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range appl...

Страница 12: ...150 KHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed abo...

Страница 13: ...s the nebuliser kit tilted at a sharp angle Make sure that the nebuliser kit is not tilted at an angle of more than 45 degrees Is the air tube connected correctly Make sure that the air tube is correc...

Страница 14: ...of wastes and recycle it responsibly to promote the sustainable reuse of material resources Household users should contact either the retailer where they purchased this product or their local governm...

Отзывы: