3) LCD Monitor Display
Monitor mode display shows the currently measured electrocardiogram signal in real
time. The monitor mode display function is used for the following
two purposes: One is
to check whether the signals from the all channels are coming out properly; the other
one is to see the status of the electrocardiogram of the patient for a
long time.
Filters for the monitor mode display are below. (Setup in system setup)
BASE LINE
On (fixed)
Electromyogram Filter (EMG)
Off (fixed)
Power Supply Filter (AC)
Option
Low Pass Filter (LPF)
40hz (fixed)
The printing speed can be selected to 12.5 mm/s, 25 mm/s, and 50 mm/s, and the
sensitivity can be selected to 2.5 mm/mV, 5 mm/mV, 10 mm/mV, 20 mm/mV, and Auto.
Display output can either be done in portrait or landscape views. If it’s in portrait
mode, press the VIEW (
) key to switch viewing mode. There are a total of 3 different
viewing modes: one displaying all 12 leads and two displays, each showing 6 leads. The
following diagram shows the display switch order with arrows indicating each press of
the VIEW key.
In the landscape mode, there are a total of two display options with either 2 6-lead
columns or 4 3-lead columns. The following diagram shows a display switch order with