1) Location Of Electrode
Attach electrodes to the patient’s body to record an electrocardiogram of twelve
standard leads [ I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 ], as shown below.
Limb electrodes are located as follows.
RL (N): Right Leg
LL (F): Left Leg
RA (R): Right Arm
LA (L): Left Arm
V1 (C1): Fourth Intercostal Space At The Right Border Of The Sternum
V2 (C2): Fourth Intercostal Space At The Left Border Of Sternum
V3 (C3): Midway Between Location V2 and V4
V4 (C4): At The Mid-Clavicular Line In The Fifth Intercostals Space
V5 (C5): At The Another Axillary Line On The Same Horizontal Level As V4
V6 (C6): At The Anterior Axillary Line On The Same Horizontal Level As V4 and V5