ecom-DP

 

   Page 23 

6. Leak Detection 

 
 
 
 
 
 
Sweep the sensor head slowly along the gas installation to be checked. As soon 
as  the  unit  nears  a  leakage,  the  display  indication  reacts  and,  by  exceeding  the 
first threshold (2 lines in display), the red LED blinks.   
 
 
 
 
 
The  blinking  red    LED  together  with  an  acoustical  signal  (disconnectible)  shows 
an increased concentration of gaseous hydrocarbons. Parallely to the  increasing 
concentration, the signal tone is getting more intensive and the red LED blinks in 
shorter intervals, so that a leakage can be accurately localized. 
 
The largest concentration registered so far is displayed with a grey bar (dragging 
announcement).  The  sensitivity  of  the  sensor  (middle  sensitivity  after  switching 
on) can be adjusted in 3 steps (see Adjustments).  
 

7. Zero Point

 

 
Indication drifts from the zero point can be corrected while pressing the 

<HOLD>

 

key  (dragging  announcement  gets  erased).  pressed  longer  than  2  seconds.  The 
change  of  the  zero  point  is  indicated  by  flashing  of  the  first  bar.  If  the 

<HOLD>

 

key  is kept pressed  longer than 2 seconds, the original  zero point gets back (no 
flashing of the first bar).

 

Substances such as silicone, hydrocarbons, leadtetraethyl, 

sulfur connections and organic phosphorus connections 

cause partially irreversible sensitivity losses  

(sensor poisoning)!

 

 

Содержание ecom-DP

Страница 1: ...Operating Instructions and ...

Страница 2: ...alues 10 5 4 Printout measurement values 10 5 5 Contrast and display backlit 10 6 Pressure check 11 7 Set clock 12 8 Internal instrument settings 8 1 Switch on off key beep 12 8 2 Pressure Auto 0 by start 13 8 3 Choose printer ecom P 13 8 4 Battery loading 13 8 5 Display damping 13 8 6 Pressure unit 14 8 7 Auto OFF function 15 8 8 Contrast 15 8 9 Light 15 9 Sensor setup 16 10 View service info 16 ...

Страница 3: ... LSG 1 Application 21 2 Instrument design 21 3 Power supply 22 4 Battery change 22 5 Operation 22 6 Leak detection 23 7 Zero point 23 8 Adjustments 24 9 Swtch off 24 10 Function control 24 11 Technical data 25 12 Maintenance tip 25 ...

Страница 4: ... 4 ecom DP 1 Application The ecom DP is designed to make easy pressure measurements and tightness checks at gas pipes and gas installations Please respect the legal safety instructions by works at gas pipes ...

Страница 5: ...nection P1 Pmax 999 9 mbar Connection P1 IR interface for IR printer Function keys current function is shown on display Connection loading unit Grafic display Cursor keys Confirmation key ON OFF key Backside Battery compartment Magnetic fixation ...

Страница 6: ...pushed directly over the quick connection If remounting the Swagelok screw connections please turn the nut firmly with your fingers and another 1 4 revolution with a wrench Connection hose 2 Connection hose 1 Connection hose ecom DP Cross piece ecom DP Swagelok screw connec tions Cross piece Compres sed air ball Shut off valve Quick connections Plug for not necessary hose Plug for not necessary co...

Страница 7: ... by beeping and flashing a signal on the LCD display If the voltage sinks approx 3 V so the ecom DP will auto matically switch off 4 2 Accu charging The accu charging procedure is monitored by the ecom DP Once connected to AC power the instrument switches on During the charging procedure a BAT TERY indicator will be shown on the LCD display Once the battery is fully charged the BATTERY indicator o...

Страница 8: ...yed Press F2 wrench symbol to call up the menu In strument settings With the cursor keys UP DOWN scroll until the menu point Set clock Press OK to activate Date and time can be set as follows 1 Select with the cursor keys RIGHT LEFT the posi tion to be changed corresponding field blinks 2 Adjust the correct numbers with the cursor keys UP DOWN 3 Press OK to confirm correct date and time settings ...

Страница 9: ... function key F1 The current seizure of the other function keys F2 F4 is shown on the display Press one time F1 to display the functions Hold Min Max and Print 5 2 Freezing of measurement value Press F2 Hold to freeze that measurement value displayed at the time the key is pressed active function Hold indication blinks Press again F2 Hold to quit the freezing and the current value will be displaye...

Страница 10: ...tween the sending diode at the front of the in strument and the receiving diode of the printer The distance between both instruments should not exceed 40 cm 1 3 foot otherwise a trouble free transmission is not warranted at any time The printout can be made with the infrared printers ecom P and HP 82240B Please choose the used print er as shown under capter 8 3 5 5 Contrast adjustment and lighting...

Страница 11: ...ompressed air ball switch off the valve 5 Start Pressure check with OK After end of stabili zation time the measurement time starts automatically 6 After end of measurement time the result is indicated on the display With F2 the result can be printed by an IR printer option 7 If the Pressure Check is selected again so the re sult can be called up with F1 as long as the ecom DP is on or a new measu...

Страница 12: ...esponding field blinks 2 Adjust the correct figures using UP DOWN 3 Confirm with OK 8 Internal instrument settings Further instrument settings can be performed in the menu Internal Press OK to access this menu 8 1 Key beep on off switch With the cursor keys UP DOWN scroll until the menu point Key beep Activate with OK With the cursor keys RIGHT LEFT modify the ad justment YES or NO Confirm with OK...

Страница 13: ... follows Scroll with the cursor keys UP DOWN until the menu point ecom P Press OK to activate Use the cursor keys RIGHT LEFT to modify the adjustment YES printer ecom P or NO printer HP 82240B Confirm the modification with OK 8 4 Battery loading The ecom DP is able to work with accus and AC power option Scroll with the cursor keys UP DOWN until the menu point Battery loading Press OK to activate U...

Страница 14: ...djustment scroll with the keys UP DOWN until the menu point Damping Press OK to activate Adjust the desired damping step using the cursor keys RIGHT LEFT Press OK to confirm 8 6 Pressure unit The measuring unit of the internal pressure sensor can be changed The following units are selectable Hectopascal hPa Centimetre water column cmH2O Millimetre mercury column mmHg Pound per square inch PSI Inch...

Страница 15: ...justed with the help of the cursor keys RIGHT LEFT Scroll with the cursor keys UP DOWN until the menu point Contrast Press OK to activate To increase the contrast use cursor key RIGHT To decrease the contrast use cursor key LEFT Confirm the modification with OK 8 9 Light The lighting of the display can be switched ON and OFF Scroll with the cursor keys UP DOWN until the menu point Light Press OK t...

Страница 16: ...sign blinks Use the UP DOWN cur sor keys to select the character that should stand on this position Starting from the current one and using the UP key the following characters are stored in the instrument in the following succession Blank sign A Z a z Ä Ü ä ß 0 9 Special characters Once the renaming procedure is complete press OK to store 10 View service information Use the cursor keys UP DOWN to ...

Страница 17: ...harge from the download area of the ecom homepage www ecom de To trans fer the data a USB cable kind becomes No 55818 and a driver free down load from www ecom de is needed The measured value rows can be stored in the xls format Excel The USB cable can be used as the power supply for the ecom DP Therefore the function Battery loading must be adjusted on no when using 1 5 V AA Mignon batteries see ...

Страница 18: ...ding on used power supply Display Graphic display Resolution 128 x 64 pixel Window approx 43 x 28 mm Operation 4 cursor keys menu monitored 1 confirmation key 4 function keys 1 ON OFF key Differential pressure measurement Pressure range 1500 hPa mbar or Pressure range 70 hPa mbar Resolution 0 1 hPa or 0 01 hPa Accuracy 3 of measurement value Internal clock Adjustment via menu and keyboard Buffer a...

Страница 19: ...rma nent several hours of measurement per day rough conditions etc shorter inter vals between checks should be selected please contact your ecom partner All ecom partners are listed under www ecom de Subject to technical changes V3 1 05 2019 ecom GmbH Am Großen Teich 2 D 58640 Iserlohn Telefon 49 0 23 71 9 45 5 Telefax 49 0 23 71 4 03 05 Internet http www ecom de eMail info ecom de ...

Страница 20: ...Page 20 ecom DP Operation Instructions ...

Страница 21: ... unit shows existing gas concentration on the display and warns with optical and acoustical signals The sensor sensibility can be adjusted so that even small leakages can be traced The robust sensor head is connected with a flexible swan neck so that also hardly accessible system parts can be reached 2 Instrument Design Please respect the legal safety instructions by works at gas pipes Swan neck D...

Страница 22: ... located on the back of the instrument and ex change the batteries respect polarity Always dispose of used batteries and ac cumulators in official recycling containers only e g in battery shops 5 Operation Switch on the ecom LSG in gas free air Hereto press the I O key until a beep is given The instrument starts a warm up phase during which the sensor is set to operation temperature running bar gr...

Страница 23: ...localized The largest concentration registered so far is displayed with a grey bar dragging announcement The sensitivity of the sensor middle sensitivity after switching on can be adjusted in 3 steps see Adjustments 7 Zero Point Indication drifts from the zero point can be corrected while pressing the HOLD key dragging announcement gets erased pressed longer than 2 seconds The change of the zero p...

Страница 24: ...lighting without acoustic signal with lighting without acoustic signal without lighting with acoustic signal with lighting with acoustic signal With the I O key the selected adjustment is taken over and the adjustment menu is left 9 Switch off Keep the I O key pressed until all bars in display are gone 10 Function control Unscrew the cover cap of the testing bottle and hold the sensor head over th...

Страница 25: ...ion moisture 20 to 80 rF Dimensions Housing 155 x 35 x 22 mm Swan neck 355 mm long Weight approx 200 g 12 Maintenance tip To secure the accuracy of your measuring instrument we recommend the annual check by an authorized ecom partner In the case of strong demand e g perma nent several hours of measurement per day rough conditions etc shorter inter vals between checks should be selected please cont...

Страница 26: ...Page 26 ecom DP Subject to technical changes 05 2019 ecom GmbH Am Großen Teich 2 D 58640 Iserlohn Telefon 02371 945 5 Telefax 02371 40305 Internet http www ecom de eMail info ecom de ...

Отзывы: