INMANUUIN148 USER MANUAL CE100 ING.ARG V.16.04.2018    3

 

This device is to be used only on healthy people having a good physical condition. 

This manual contains all the relevant information on the appropriate care and handling of the device while 
on operation, as well as information about its accessories. 

Please read carefully this manual in whole, before using the device for the first time. 

 

 This  symbol  shall  be  used  in  the  manual  to  indicate  information  that  includes  important  warnings  or 
precautions. 

 

Should you have any doubt regarding the contents of the manual herein, or the operation and safety of the 
device, please contact your local distributor or the nearest ECLERIS office. 
 

2.

 

 

WARNING AND ADVICE 

 

Please ensure that you turn off the device once treatment is completed (ON/OFF switch). 

 

It is advisable not to connect the patient to the applied parts before turning the device on. 

 

Do  not  conduct  the  treatment  on  areas  where  the  skin  is  damaged,  sick  or  displaying  unhealed 
wounds. 

 

Do not apply this treatment to people with pacemakers. 

 

Do not apply this treatment to people allergic to the medical grade surgical steel of the needle. 

 

Do not apply this treatment to pregnant women. 

 

Do not apply this treatment to people with epilepsy or heart conditions. 

 

Do  not  apply  this  treatment  to  people  incapable  of  expressing  or  disclosing  an  abnormal  feel 
(children, elderly people) 

 

 The device operator shall ensure that the above described conditions are not present, before proceeding to 
apply this treatment to the patient. 

 

 

Keep the device outside the reach of children. 

 

Do not expose the device to dust, humidity or temperatures lower than 15ºC (59ºF) or higher than 
40ºC (104°F).  

 

This device shall only be used by personnel informed of the above stated warnings. 

 

After each use of the device, turn it off and unplug the power supply cable. 

 

When the device is not in use, protect it against unauthorized use. 

 

Do  not  deform,  bend,  step  on  or  tie  the  power  supply  cable  or  any  other  cable  connected  to  the 
device. 

 

Replace said cables when the first signs of wear appear and use only original parts. 

 

Do  not  obstruct  the  aeration  outlets.  This  aeration  must  be  placed  in  such  a  way  that  air  flows 
smoothly. 

 

Do  not  place  containers  with  liquids  on  top  of  the  device  and  keep  it  away  from  the  rain  and 
sunlight. 

 

Any use other than the one intended exempts the manufacturer from any and all liability. 

 

The  device  must  not  be  used  connected  to  an  electrical  power  supply  network  in  breach  of  the 
regulations in force in the country or to a not-grounded electrical network. 

 

Содержание SPIDER

Страница 1: ...INMANUUIN148 USER MANUAL CE100 ING ARG V 16 04 2018 SPIDER VASCULAR TERMOCOAGULATOR...

Страница 2: ...3200 Web www ecleris com E Mail info ecleris com Directora T cnica Farm Doris In s Rasi Sanchez MN 12 856 Autorizado por la ANMAT PM 1120 36 Condici n de venta Venta exclusiva a profesionales e instit...

Страница 3: ...System configuration 9 9 4 Body treatments 10 9 5 Face treatment 11 9 6 Customized programs 11 9 7 Placing the needle for vascular lesions treatments 12 10 TREATMENTS 13 10 1 General precautions 13 1...

Страница 4: ...insulating material which in no case will affect the surrounding tissues the thermo coagulator provides an accurate scope of operation but no side effects It consists in a main radiofrequency unit th...

Страница 5: ...nant women Do not apply this treatment to people with epilepsy or heart conditions Do not apply this treatment to people incapable of expressing or disclosing an abnormal feel children elderly people...

Страница 6: ...in needle that is pressed almost perpendicular to and very superficially on the skin where the small vessel requiring treatment is located and which sends a highly efficient and accurate regulated the...

Страница 7: ...age 100 240 V N of phases 1 single phase Frequency 50 60 Hz Protection level against ingress of water and particles IPX0 Pedal IPX1 Consumption 100VA Fuse 2 x T2AL250V Cooling system Air forced convec...

Страница 8: ...Read the related documentation Warning Alternate current Protect from water This side up Environmental transport conditions Fragile Manufacturers information Name Address Type of applied BF equipment...

Страница 9: ...provided in this section 8 1 Unpacking 1 Remove the machine from the box 2 Check whether damage has occurred as a result of transportation 3 Ensure that everything has been delivered 4 In case of mis...

Страница 10: ...department of your local distributor or the nearest ECLERIS office prior to its use 9 2 Operation This system must be used only for medical paramedical or health professionals This device was not des...

Страница 11: ...alarm that triggers once the time stated in this control is up Every time treatment is started i e pushing the pedal for the first time within the selected program the TIMER will start the countdown...

Страница 12: ...nal so decides During treatment the values can be changed in the preset programs but such changes will not be saved You can change the POWER PULSE duration and DELAY This last feature represents the t...

Страница 13: ...any of these options will display a new screen containing the following 3 parameters POWER PULSE duration and DELAY described above in the Body Treatment section the values of which have been preset a...

Страница 14: ...ation value 9 7 Placing the needle for vascular lesions treatments SPIDER can be used with 2 needle types the smallest diameter 0 003 0 075 mm needle is used for very small blood vessels and the large...

Страница 15: ...ve or can it be applied on objects or animals 10 3 Before starting treament Before starting any treatment please read carefully the instructions below Disinfect the area to be treated Place the needle...

Страница 16: ...ro edema and erythema may be observed on the treated areas They will go away within a couple of days as the treatment develops You should warn the patient so that he she is aware of it A regenerating...

Страница 17: ...tion an excess application on the treated area Such pigmentation is often temporary and normally subsides Given a lack of desired results the following must be checked That the varicose vein size is a...

Страница 18: ...cision and safe technology No need of previous preparation by the patient Fast and user friendly technology Immediate results Applicable all year round For all types of skin It can be applied in any p...

Страница 19: ...ot use detergents solvents or any other abrasive agent Do not send to the autoclave any part or component of this device that are not specifically marked as autoclavable Do no sterilize or immerse the...

Страница 20: ...ty you may encounter PROBLEM PROBABLE CAUSE ACTION The device does not turn on It is not connected to the electrical network Connect the power supply cable to the electrical network The power supply c...

Страница 21: ...vice during thunderstorms After the use of the device turn it off and disconnect the power cable SPIDER is grounded through a conductor of the power supply cable This essential grounding is provided f...

Страница 22: ...c waste Research the local rules and laws for proper disposal It is the user responsibility to deliver the system to the proper location for disposal or recycle of electrical and electronic device or...

Страница 23: ...acturer We recommend using the following procedure to speed up the repair service and replacement of products under warranty 1 Contact the technical service department of your local distributor or the...

Страница 24: ...voice ECLERIS shall not be liable for any damage caused to by SPIDER due to negligence incorrect use cleaning or storage Any repairs made by people not authorized by ECLERIS shall void this warranty T...

Страница 25: ...t of this manual shall be reproduced by any means photocopy microfilm or other methods without the previous written authorization by ECLERIS likewise no part shall be reproduced multiplied or publishe...

Отзывы: