INMANUUIN148 USER MANUAL CE100 ING.ARG V.16.04.2018
SPIDER
VASCULAR TERMOCOAGULATOR
Страница 1: ...INMANUUIN148 USER MANUAL CE100 ING ARG V 16 04 2018 SPIDER VASCULAR TERMOCOAGULATOR...
Страница 2: ...3200 Web www ecleris com E Mail info ecleris com Directora T cnica Farm Doris In s Rasi Sanchez MN 12 856 Autorizado por la ANMAT PM 1120 36 Condici n de venta Venta exclusiva a profesionales e instit...
Страница 3: ...System configuration 9 9 4 Body treatments 10 9 5 Face treatment 11 9 6 Customized programs 11 9 7 Placing the needle for vascular lesions treatments 12 10 TREATMENTS 13 10 1 General precautions 13 1...
Страница 4: ...insulating material which in no case will affect the surrounding tissues the thermo coagulator provides an accurate scope of operation but no side effects It consists in a main radiofrequency unit th...
Страница 5: ...nant women Do not apply this treatment to people with epilepsy or heart conditions Do not apply this treatment to people incapable of expressing or disclosing an abnormal feel children elderly people...
Страница 6: ...in needle that is pressed almost perpendicular to and very superficially on the skin where the small vessel requiring treatment is located and which sends a highly efficient and accurate regulated the...
Страница 7: ...age 100 240 V N of phases 1 single phase Frequency 50 60 Hz Protection level against ingress of water and particles IPX0 Pedal IPX1 Consumption 100VA Fuse 2 x T2AL250V Cooling system Air forced convec...
Страница 8: ...Read the related documentation Warning Alternate current Protect from water This side up Environmental transport conditions Fragile Manufacturers information Name Address Type of applied BF equipment...
Страница 9: ...provided in this section 8 1 Unpacking 1 Remove the machine from the box 2 Check whether damage has occurred as a result of transportation 3 Ensure that everything has been delivered 4 In case of mis...
Страница 10: ...department of your local distributor or the nearest ECLERIS office prior to its use 9 2 Operation This system must be used only for medical paramedical or health professionals This device was not des...
Страница 11: ...alarm that triggers once the time stated in this control is up Every time treatment is started i e pushing the pedal for the first time within the selected program the TIMER will start the countdown...
Страница 12: ...nal so decides During treatment the values can be changed in the preset programs but such changes will not be saved You can change the POWER PULSE duration and DELAY This last feature represents the t...
Страница 13: ...any of these options will display a new screen containing the following 3 parameters POWER PULSE duration and DELAY described above in the Body Treatment section the values of which have been preset a...
Страница 14: ...ation value 9 7 Placing the needle for vascular lesions treatments SPIDER can be used with 2 needle types the smallest diameter 0 003 0 075 mm needle is used for very small blood vessels and the large...
Страница 15: ...ve or can it be applied on objects or animals 10 3 Before starting treament Before starting any treatment please read carefully the instructions below Disinfect the area to be treated Place the needle...
Страница 16: ...ro edema and erythema may be observed on the treated areas They will go away within a couple of days as the treatment develops You should warn the patient so that he she is aware of it A regenerating...
Страница 17: ...tion an excess application on the treated area Such pigmentation is often temporary and normally subsides Given a lack of desired results the following must be checked That the varicose vein size is a...
Страница 18: ...cision and safe technology No need of previous preparation by the patient Fast and user friendly technology Immediate results Applicable all year round For all types of skin It can be applied in any p...
Страница 19: ...ot use detergents solvents or any other abrasive agent Do not send to the autoclave any part or component of this device that are not specifically marked as autoclavable Do no sterilize or immerse the...
Страница 20: ...ty you may encounter PROBLEM PROBABLE CAUSE ACTION The device does not turn on It is not connected to the electrical network Connect the power supply cable to the electrical network The power supply c...
Страница 21: ...vice during thunderstorms After the use of the device turn it off and disconnect the power cable SPIDER is grounded through a conductor of the power supply cable This essential grounding is provided f...
Страница 22: ...c waste Research the local rules and laws for proper disposal It is the user responsibility to deliver the system to the proper location for disposal or recycle of electrical and electronic device or...
Страница 23: ...acturer We recommend using the following procedure to speed up the repair service and replacement of products under warranty 1 Contact the technical service department of your local distributor or the...
Страница 24: ...voice ECLERIS shall not be liable for any damage caused to by SPIDER due to negligence incorrect use cleaning or storage Any repairs made by people not authorized by ECLERIS shall void this warranty T...
Страница 25: ...t of this manual shall be reproduced by any means photocopy microfilm or other methods without the previous written authorization by ECLERIS likewise no part shall be reproduced multiplied or publishe...