4.  INSTALLATION 

The installation of the HD Colposcope does not require skilled or specialized staff. In addition to 
the tools supplied, the following tools are required: one 13 mm wrench, one 27 mm wrench or an 
adjustable one. 

In  order  to  avoid  accidents  caused  by  the  manipulation  of  the  different  parts  of  the 
device,  it  is  recommended  that  two  people  perform  the  installation,  following  these 
instructions carefully. 

Do not install the device near sources of heat such as heaters. 

Install the device on firm and horizontal surfaces with an inclination of no more than 5°. 

Avoid  exposing  the  device  and  its  accessories  to  direct  sunlight,  dust,  or  excessive 
vibrations or bumps. 

Remove the device’s parts from their packaging. 

Check that the device is complete.  

Check that the device has not been damaged by the transportation. 

If possible, keep the original packaging in case you have to transport the device again. 

4.1 MOUNTING THE BASE 

1) 

Place  the  screw  (bolt)  in  the  base’s  central  orifice.  This  screw  has  a  27  mm  hexagonal 
head and does not require a washer. 

 

 

 

 

 

 

2)  Put the column and screw the screw on same. Use a 27 

mm wrench to fix the screw to the column. 

 

 

 

 

 

Once the column is conveniently fixed to the base, you may turn the base on the four wheels and 
proceed with the next step. 

Содержание C100HD

Страница 1: ...without previous written authorization from ECLERIS Neither can it be reproduced multiplied or published through any electronic data processing form The information contained in this manual may be mod...

Страница 2: ...BASE 6 4 2 MOUNTING THE SET TO THE GANTRY 7 4 3 MOUNTING THE CONSOLE 7 4 4 THE GANTRY S ADJUSTMENTS 8 5 CONTROLS 9 5 1 HEAD 9 5 2 THE CONSOLE S FRONT 9 5 3 THE CONSOLE S BACK PANEL 10 6 OPERATION 11...

Страница 3: ...ould be used in offices where the electrical installation is grounded pursuant to the relevant rulings for health facilities The HD VIDEO COLPOSCOPE should be connected to the mains electricity throug...

Страница 4: ...vice during electrical storms Do not introduce objects into the orifices and do not place containers with fluids on the device Disconnect the power supply cable from the outlet when the device is not...

Страница 5: ...hic studies in vivo and measuring the size of the lesions with capacity for diagnosing a multiplicity of them The Video Colposcopy study allows identifying the subtle characteristics of the epithelium...

Страница 6: ...irm and horizontal surfaces with an inclination of no more than 5 Avoid exposing the device and its accessories to direct sunlight dust or excessive vibrations or bumps Remove the device s parts from...

Страница 7: ...light source below the forearm matching the supports with the orifices of the console s front face 3 Move the console upwards so that the supports are introduced into the console s orifices 4 Move th...

Страница 8: ...mechanical brakes to regulate the arm and the forearm s different positons as well as regulating the head s height so as to ensure a stable focus and the operator s comfort The head also has a brake...

Страница 9: ...may be used with the light beam so as to have a special lighting on the tissues This increases the contrast of the vascular beds and provides a better histological differentiation on the basis of the...

Страница 10: ...back panel hosts the power supply connector with fuse holder the equipotential ground terminal which should be duly connected to the equipotential ground node of the medical facility and two HD video...

Страница 11: ...y with the light off the user may control it manually using the joystick F F White balance The white balance is set from the factory and it is stored in the device s non volatile memory However if it...

Страница 12: ...NU In order to access to the advanced menu press any key on the device s head while it is turning on while the three head s lights are flickering for approximately 10 seconds Use the joystick s comman...

Страница 13: ...ions Revision of the electrical safety parameters 7 1 CLEANING Before cleaning the HD colposcope disconnect the cord from the mains electricity Do not sterilize or submerge the device or its accessori...

Страница 14: ...cover to replace the damaged fuse 3 Remove the damaged fuse and replace it with another one 4 Use the fuses pursuant to the marks on the device s labels 5 Introduce the socket again in the same positi...

Страница 15: ...240 300 mm minimum magnification Field of view 90x50 mm 300 mm with minimum magnification Magnification Zoom levels depend on the monitor size in the following table some of the possible monitor size...

Страница 16: ...eo cable is disconnected Check the connection between the video cable and the console s front panel The head does not emit light The video connector is disconnected Check the connection between the vi...

Страница 17: ...tion for the distance from the target Decrease the magnification level See zoom level The central zone of the image has multiple focal distances Deactivate the autofocus and perform the focus adjustme...

Страница 18: ...ect in the manufacture of same Therefore the appropriate service under the applicable conditions of use indicated herein is guaranteed The warranty period is of one 1 year effective as of the date of...