  The  device  is  not  apt  to  be  used  in  presence  of  an  inflammable  anesthetic  mix  with  air, 

oxygen or nitrous oxide 

Be careful when moving the head up and down, since there is risk to trap the hand or 
the  fingers  between  the  device’s  arm  and  the  head’s  base.  Take  the  handwheel  to 
regulate the head’s height with one hand and keep the forearm firm with the other hand 
in a safe position, away from the area with risk to smashing the fingers. 

 

  Do not look straight to the light output, since the head has a very brilliant light output. 

  While  moving  the  device  (during  the  installation  and  when  the  device  is  positioned  for 

performing a study) check that the light is off so as to avoid illuminating the face straightly. 

  Do not use the device during electrical storms. 

  Do  not  introduce  objects  into  the  orifices  and  do  not  place  containers  with  fluids  on  the 

device. 

  Disconnect the power supply cable from the outlet when the device is not in use. 

  In case of interference with other electro medical or electrical devices with a low immunity 

level,  it  is  recommended  to  increase  the  distance  between  the  device  and  the  interfered 
devices so as to reduce those effects. 

  Do  not  open  the  device.  Maintenance  and  repairs  should  be  made  by  ECLERIS  or  by  a 

Company authorized by same. ECLERIS shall not be liable for the consequence of repairs 
of maintenance performed by unauthorized persons. 

  This instructions manual should stay near the device for future consult. 

 

If, the consequence of: 

  The inappropriate use of the device, 

  Not using the instructions manual, 

  Misuse or inappropriate maintenance of the device, 

  Repairs or manipulation performed by unauthorized staff,  

  Continuing using the device in case of malfunction or unstable performance,  

Содержание C100HD

Страница 1: ...without previous written authorization from ECLERIS Neither can it be reproduced multiplied or published through any electronic data processing form The information contained in this manual may be mod...

Страница 2: ...BASE 6 4 2 MOUNTING THE SET TO THE GANTRY 7 4 3 MOUNTING THE CONSOLE 7 4 4 THE GANTRY S ADJUSTMENTS 8 5 CONTROLS 9 5 1 HEAD 9 5 2 THE CONSOLE S FRONT 9 5 3 THE CONSOLE S BACK PANEL 10 6 OPERATION 11...

Страница 3: ...ould be used in offices where the electrical installation is grounded pursuant to the relevant rulings for health facilities The HD VIDEO COLPOSCOPE should be connected to the mains electricity throug...

Страница 4: ...vice during electrical storms Do not introduce objects into the orifices and do not place containers with fluids on the device Disconnect the power supply cable from the outlet when the device is not...

Страница 5: ...hic studies in vivo and measuring the size of the lesions with capacity for diagnosing a multiplicity of them The Video Colposcopy study allows identifying the subtle characteristics of the epithelium...

Страница 6: ...irm and horizontal surfaces with an inclination of no more than 5 Avoid exposing the device and its accessories to direct sunlight dust or excessive vibrations or bumps Remove the device s parts from...

Страница 7: ...light source below the forearm matching the supports with the orifices of the console s front face 3 Move the console upwards so that the supports are introduced into the console s orifices 4 Move th...

Страница 8: ...mechanical brakes to regulate the arm and the forearm s different positons as well as regulating the head s height so as to ensure a stable focus and the operator s comfort The head also has a brake...

Страница 9: ...may be used with the light beam so as to have a special lighting on the tissues This increases the contrast of the vascular beds and provides a better histological differentiation on the basis of the...

Страница 10: ...back panel hosts the power supply connector with fuse holder the equipotential ground terminal which should be duly connected to the equipotential ground node of the medical facility and two HD video...

Страница 11: ...y with the light off the user may control it manually using the joystick F F White balance The white balance is set from the factory and it is stored in the device s non volatile memory However if it...

Страница 12: ...NU In order to access to the advanced menu press any key on the device s head while it is turning on while the three head s lights are flickering for approximately 10 seconds Use the joystick s comman...

Страница 13: ...ions Revision of the electrical safety parameters 7 1 CLEANING Before cleaning the HD colposcope disconnect the cord from the mains electricity Do not sterilize or submerge the device or its accessori...

Страница 14: ...cover to replace the damaged fuse 3 Remove the damaged fuse and replace it with another one 4 Use the fuses pursuant to the marks on the device s labels 5 Introduce the socket again in the same positi...

Страница 15: ...240 300 mm minimum magnification Field of view 90x50 mm 300 mm with minimum magnification Magnification Zoom levels depend on the monitor size in the following table some of the possible monitor size...

Страница 16: ...eo cable is disconnected Check the connection between the video cable and the console s front panel The head does not emit light The video connector is disconnected Check the connection between the vi...

Страница 17: ...tion for the distance from the target Decrease the magnification level See zoom level The central zone of the image has multiple focal distances Deactivate the autofocus and perform the focus adjustme...

Страница 18: ...ect in the manufacture of same Therefore the appropriate service under the applicable conditions of use indicated herein is guaranteed The warranty period is of one 1 year effective as of the date of...

Отзывы: