Chapter 4. Handheld mode measurement
4.2 Audiometry
The “Settings” button opens a window allowing you to
set the level of the masking noise as well as the starting
power of the automatic protocols. Click “OK” to con-
firm
After selecting the ear (right or left), the
«Start»
button became available. Click on it to start the measurement.
Measurement
The
Tone audiometry
window open. The
AUDI-
O
SMART
device will automatically scan the pre config-
ured frequencies and increase or decrease the stimuli
power according to the patient answers.
The patient simply has to click on “OK” or on the re-
sponse button when he hears the sound.
If the click has been taken into account, the button turns
orange.
Once the acquisition protocol ends, the curve is constructed. You have the choice between saving the curve with the
“Save” button or delete it by leaving the window with the back cross.
4.2.1.2
Automatic doctor mode
Automatic doctor mode allows automatic transitions of powers and frequencies. During the test, the device display
the power and the frequency of the current stimuli. This mode allow the doctor to easily make a test while checking the
patient responses.
Measurement settings
Click on “
Settings
” then on “
Frequencies selection
” to preconfigure the frequencies to run during the test. Once they
are chosen, click on “
OK
” to validate.
To obtain more information about curve consulting, please refer to the paragraph
The saved data can be found in the “
Consultation
” patient menu.
All frequencies can be selected, however, the maximum and minimum stimulation frequencies may be
limited at the time of the test depending on the characteristics of the stimulator.