10
EN
Echö
MAINTENANCE
The transducer is fragile and must be handled with care. Please check that the Echö
does not have visible signs of damage that may affect user safety ormonitoring capability
before use.
• Carefully inspect the equipment for physical damage.
• Examine the case, connectors and accessories making sure every part is clean and is
not cracked or damaged in any way.
• After the visual inspection, if the Echö and/or its accessories are damaged,
replacement is needed.
• Before cleaning, turn off the Echö
E
, disconnect the earphones
A
and the
recording cable
C
and remove the battery.
• If gel is used, clean it from the transducer probe
B
with a soft cloth. You can
disinfect the probe by wiping it with 70% ethanol or alcohol wipes and then drying
with a soft dry cloth.
• Do not use abrasives, solvents, bleach, steam or high temperature cleaning
processes to sterilize the Echö as any of these can damage it.
• Do not immerse in water or other liquids to clean the Echö.
• Do not disassemble the Echö.
• Checking the battery level
NOTE - The Echö requires no calibration.
TROUBLESHOOTING
What appears to be a malfunction may not always be serious. If your Echö does not
perform as expected, consult the table below to see if the problems can be corrected.
If you are unable to resolve the problem, contact bblüv Customer Service at
1-855-808-2636 or by email at [email protected]
NOTE: Make sure to always test the Echö with your own heartbeat first to make sure it is working correctly