32
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
Ebb is intended for use in the electromagnetic environment specified below. The customer or the user of Ebb should assure that it is used in
such an environment.
IMMUNITY test IEC 60601 TEST LEVEL
Compliance level
Electromagnetic environment - guidance
IMMUNITY
to proximity
fields from RF
wireless
communications
equipment
MHz – Modulation –Field
Strength
385 - 18 Hz - 27 V/m
450 - 18 Hz - 28 V/m
710 - 217 Hz - 9 V/m
745 - 217 Hz - 9 V/m
780 - 217 Hz - 9 V/m
810 - 18 Hz - 28 V/m
870 - 18 Hz - 28 V/m
930 - 18 Hz - 28 V/m
1720 - 217 Hz - 28 V/m
1845 - 217 Hz - 28 V/m
1970 - 217 Hz - 28 V/m
2450 - 217 Hz - 28 V/m
5240 - 217 Hz - 9 V/m
5500 - 217 Hz - 9 V/m
5785 - 217 Hz - 9 V/m
MHz – Modulation – Field
Strength
385 - 18 Hz - 27 V/m
450 - 18 Hz - 28 V/m
710 - 217 Hz - 9 V/m
745 - 217 Hz - 9 V/m
780 - 217 Hz - 9 V/m
810 - 18 Hz - 28 V/m
870 - 18 Hz - 28 V/m
930 - 18 Hz - 28 V/m
1720 - 217 Hz - 28 V/m
1845 - 217 Hz - 28 V/m
1970 - 217 Hz - 28 V/m
2450 - 217 Hz - 28 V/m
5240 - 217 Hz - 9 V/m
5500 - 217 Hz - 9 V/m
5785 - 217 Hz - 9 V/m
Portable and mobile RF communications equipment should
be used no closer to any part of Ebb, including cables, than
the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
E = [6/d] √P
d = [6/E] √P
where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer, d is the
recommended separation distance in meters (m), and E is the
field strength in V/m. Field strengths from fixed RF transmitters,
as determined by an electromagnetic site survey, should be
less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with
the following symbol:
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
• Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in which Ebb is used exceeds the applicable RF compliance
level above, Ebb should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-
orienting or relocating Ebb.