4
INM MTL 130-0186 Rev 8
3
INSTALLATION
3.1 Unpacking and visual checking
Take all normal precautions when opening packages. In particular, avoid the use of long bladed
cutters. Check for any sign of damage. Check that all pipe connections have captive seal nuts -
search packing if any are missing.
3.2 Calibration overview
The instrument should be installed where free air ventilation around the whole case is
provided. The ambient temperature should not exceed 45°C. Dust and dirt should be kept to a
minimum. Ambient air must contain a normal Oxygen level.
WARNING
The case should not be exposed to water jets or drips.
Figure 1
- Instrument dimensions – 19” rack mounting 3U high
3.3 Sample
WARNING
It is important that no flammable mixtures are allowed to come into contact
with the measuring cell as this may cause ignition of the gas. Gases
containing halogens, sulphur or silicon, must be avoided. It is the responsibility
of the user to ensure that any advice offerred can be safely applied to their
particular situation
CAUTION
THE DEW POINT OF THE SAMPLE MUST ALWAYS BE LESS THAN THE AMBIENT
TEMPERATURE TO AVOID THE RISK OF LIQUID DROPLETS FORMING IN THE
MEASURING CELL