Guidance and manufacturer’s declaration – electromagnetic emissions
This device is intended for use in the electromagnetic environment specified below. The customer or the user of this device should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissionsCISPR 11
Group 1
This device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic equipment.
RF emissionsCISPR 11
Class B
This device is suitable for use in all establishments, including domestic establishments and those directly
connected to the public low-voltage power supply network that supplies buildings used for domestic
purposes.
Harmonic emissionsIEC 61000-3-2
Not applicable
Voltage fluctuations/flicker emissionsIEC 61000-3-3
Not applicable
Guidance and manufacturer’s declaration – electromagnetic immunity
This device is intended for use in the electromagnetic environment specified below. The customer or the user of this device should assure that it is used in such an environment.
Immunity test
IEC 60601test level
Compliance level
Electromagnetic environment –guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30 %.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels characteristic of a typical
location in a typical commercial or hospital environment.
NOTE
U